This website uses cookies. By using the site you are agreeing to our Privacy Policy.

Endotoxin Detection ReagentsPYROSTAR™ ES-F/Plate

  • Turbidimetric Technique
  • Endotoxin Specific
  • FDA licensed
item.jpg

PYROSTAR™ ES-F / Plate is a specific Turbidimetric lysate reagent specifically for 96-well microplates.
By using an absorbance microplate reader, simultaneous measurement of multiple samples is easily made possible.
The quantitative range is wide, from 0.01 to 10 EU / mL, and it can be applied to water used to manufacture pharmaceuticals, as well as to the testing of pharmaceuticals that require dilution.

Features

  • It is possible to specifically detect endotoxin without it being affected by (1 → 3) - β-D - Glucan in the sample.
  • Regarding U.S. exports, it can also be applied to the final product testing.
  • Since it can be measured with a sample volume of 50 μL per well, even a precious sample can be measured in a small amount.
  • Stable endotoxin measurement is possible, due to the minor pH impact of the measurement sample on testing.

PYROSTAR™ ES-F/PLATE TEST

Multi test
Catalog No. Product Name Quantitative Range
(EU/mL)
Quantity
543-10331
with CSE
PYROSTAR™ ES-F/PLATE TEST 2 mL x 4 vials 0.01-10 160 tests
540-10341
with CSE
PYROSTAR™ ES-F/PLATE TEST 5.2 mL x 4 vials 0.01-10 400 tests

CSE: 1 vial (500ng/vial)

Kit: ES-F/Plate(543-10331) reagent for 2 mL (40 tests with microplate reader) x 4 vials; Control Standard Endotoxin (CSE) x 1 vial
Kit: ES-F/Plate(540-10341) reagent for 5.2 mL (100 tests with microplate reader) x 4 vials; Control Standard Endotoxin (CSE) x 1 vial
LAL reagent solution can be stored below −15 ±5℃ for up to 2 weeks.

PYROSTAR™ ES-F/PLATE TEST (Bulk)

2mL
Catalog No. Manufacturer Code No. Product Name Quantity
548-10401 WPEPM-20015 PYROSTAR™ ES-F/PLATE TEST 2 mL x 100 vials 4000 tests

Kit: ES-F/Plate(548-10401) reagent for 2 mL x 100 vials
LAL reagent solution can be stored below −15 ±5℃ for up to 2 weeks

standard curve information

f1.jpg

Regression equation: log(Tg) = ー0.260424 log(C) +1.10351
Correlation coefficient: ー0.9943

Endotoxin concentration
(EU/mL)
Gelation time (min.)
0.0100 46.0
46.7
43.3
0.1000 21.2
21.5
21.2
1.000 12.0
12.1
11.9
10.000 7.4
7.4
7.5

Recovery rate of spiked endotoxins with the Turbidimetric technique

f2.jpg
f3.jpg

Technical Videos

LAL technical videos

Topics

  • LAL technical videos
  • banner_fix_t3.png

Contact

For inquiries please click here

CONTACT