LBIS (TM) Mouse IL-6 ELISA Kit

for Immunochemistry
Manufacturer :
FUJIFILM Wako Pure Chemical Corporation
Storage Condition :
Keep at 2-10 degrees C.
GHS :
  • Structural Formula
  • Label
  • Packing
SDS
Comparison
Product Number
Package Size
Price
Availability
Certificate of Analysis
Purchase
Distributor
299-96001
Barcode No
4548995099433
96Tests
List Price
858.00 USD

In stock in Japan

Document

SDS
Product Specification Sheet
Package Insert
Spectral Data
Certificate of Analysis
Calibration Certificate
Analytical Charts

Kit component

96 tests

Antibody-coated Plate 1 plate
Mouse IL-6 Standard (Freeze-dried) 2 vials
Buffer 30 mL
Biotin-conjugated Antibody Solution 100 μL
Peroxidase-conjugated Streptavidin Solution 100 μL
TMB Solution 12 mL
Buffer for Standard and Sample 30 mL
Stop Solution 12 mL
Wash Solution (10x) 100 mL
Plate seal 4 sheets

Product Overview

04545179_img01.jpg

IL-6 (Interleukin-6) is a secreted glycoprotein of 187 amino acids (in case of mice) and it was discovered to be a cytokine that differentiates B cells into antibody-producing cells. It is also known to play an important role in various physiological processes such as immune response, inflammatory response, hematopoiesis, and the proliferation and differentiation of nervous system cells.

After being produced by T cells, monocytes, fibroblasts, endothelial cells, keratinocytes, etc., it binds to the IL-6 receptor, which consists of two chains of IL-6Rα and gp130, and signals are transduced. There is also a study that suggested that a possible correlation exists between IL-6 levels and disease activity in rheumatoid arthritis, and it is also attracting attention in the field of autoimmune diseases such as rheumatoid arthritis and inflammatory diseases.

This kit is a sandwich ELISA system for quantitative measurement of mouse IL-6. It can measure IL-6 in mouse serum, plasma and culture supernatant, specifically and with high sensitively.

Features

  • Designed for specific and highly sensitive measurement of IL-6 in mouse serum, EDTA plasma, and cell culture supernatant
  • Requires a total measurement duration of 3 hours and 50 minutes
  • The standard is a recombinant protein produced by E. coli and is not subject to the Cartagena Protocol
  • The standard is divided into two vials to allow for multiple measurements

Kit Performance

Calibration curve range 2.05~500 pg/mL
Assay target IL-6 (Interleukin-6)
Analysis sample Mouse serum/plasma (EDTA)/cell culture supernatant
Sample volume 25 μL (2-fold dilution)
Measurement duration 3 hours 50 minutes
Detection method Chromogenic (Primary wavelength 450 nm / reference wavelength 620 nm)

Example of Calibration Curve

04545179_img02.png

Data

Precision of assay (within assay variation)

Two serum samples, each at a different concentration, were measured in quintuplicate using this kit to confirm reproducibility.

n\ID ID1 (pg/mL) ID2 (pg/mL)
1 281 18.7
2 285 19.1
3 281 18.7
4 282 18.7
5 291 19.6
mean 284 19.0
SD 4.18 0.390
CV(%) 1.47 2.06
[Result]
The CV (%) was 1.47% and 2.06%, indicating good reproducibility.

Reproducibility (between assay variation)

Three serum samples, each at a different concentration, were measured in triplicate over four consecutive days using this kit.

Day\ID ID3 (pg/mL) ID4 (pg/mL) ID5 (pg/mL)
1 246 84 15.1
2 230 78.8 13.9
3 254 86.2 15.3
4 244 84.9 14.4
mean 244 83.5 14.7
SD 9.90 3.23 0.658
CV(%) 4.07 3.87 4.49
[Result]
The CV (%) ranged from 3.87% to 4.49%, indicating good reproducibility.

Dilution linearity test

Samples of serum, EDTA plasma, D-MEM medium, and 10% FCS/D-MEM medium, each at two different concentrations, were subjected to a 2-fold serial dilution with sample buffer and measured to confirm linearity (measured in duplicates).

04545179_img03.png
04545179_img04.png
04545179_img05.png
04545179_img06.png
[Result]
Good linearity was confirmed for all samples within the calibration curve range.

Spiked recovery test

Standard solutions at four concentrations were added to serum, EDTA plasma, D-MEM medium, and 10% FCS/D-MEM medium samples. Next, spike-recovery tests were conducted, with each measurement performed in duplicate.

Amount spiked (pg/mL) Measured value (pg/mL) Recovery volume (pg/mL) Recovery rate (%)
Serum - 4.8 - -
22.8 26.9 22.1 96.9
46.0 49.0 44.2 96.0
96.4 94.9 90.1 93.4
197.0 193 188 95.5
Average 95.5
Plasma (EDTA) - 4.4 - -
28.1 30.6 26.2 93.0
56.0 55.0 50.5 90.2
116 107 102 88.4
227 206 202 88.8
Average 90.1
D-MEM - 0.0 - -
23.0 24.1 24.1 105.0
48.3 47.7 47.7 98.7
97.5 93 93.0 95.3
194.4 205 205.1 105.5
Average 101.1
10% FCS /D-MEM - 0.0 - -
28.1 27.6 27.6 98.1
56.0 60.2 60.2 107.5
115.6 121 120.7 104.4
227.3 237 237.2 104.3
Average 103.6
[Result]
The recovery rates ranged from 88.4% to 107.5%, indicating good performance.

Cross-reactivity with similar proteins

Proteins similar to mouse IL-6, including IL-6 of other species and other interleukins, were tested to determine if they could be detected.

Proteins Amount spiked (pg/mL) Measured value (pg/mL)
Human IL-6 500 0.034
Rat IL-6 500 0.214
Mouse IL-4 1,000 0.01
Mouse IL-7 1,000 -
Mouse gp130 1,000 -
Mouse IL-6Rα 1,000 -

-: The values are outside the standard curve.

[Result]
Very low reactivity with similar proteins was confirmed.

Correlation with conventional product

For mouse serum samples, measurements were conducted using this kit and competitor product A. The measurement values obtained using both products were then compared.

04545179_img07.png
[Result]
A good correlation was confirmed.

Detection Performance Comparison

Measurements were conducted on mouse serum samples using this kit and competitor product A to compare their detectability.

Sample No. Fujifilm Wako (pg/mL) Company A (pg/mL)
1 1.71 Not detectable
2 6.50 5.52
3 6.39 9.02
4 7.01 7.67
5 9.67 7.53
Proportion of samples
successfully measured
4/5 (80%) 1/5 (20%)
Samples measured within the calibration range are indicated in red; samples outside the range are noted as reference values.
[Result]
This kit demonstrated high detection sensitivity, successfully detecting four out of five samples within the calibration curve range.

FAQ

About sample

How should samples be stored?
For long-term storage of samples, please store below –35℃ until assay. Before starting the assay, thoroughly shake thawed samples. Do not repeat freeze-and-thaw cycles.
Can I measure heparin plasma or citric acid plasma as a sample?
Please use only EDTA as anticoagulant.
How should samples with turbidity or insoluble matter be handled?
Centrifuge the sample to remove turbidity or insoluble matter where necessary before use in the assay.

About the kit

How do I calibrate the measured values?
The measured value obtained by this kit can be converted to the unit concentration based on NIBSC/WHO standard mouse IL-6 (Code: 93/730) by multiplying with the conversion factor specified in a separate sheet according to the following calculation formula.
NIBSC/WHO(93/730) unit concentration (U/mL)=Conversion factor x Measured mouse IL-6 value (pg/mL) *The conversion factor is different depending on the lot. Please see the specified conversion factor in the annex document.
Is it possible to purchase the antibodies provided in this kit?
These antibodies are not sold separately.
Can I divide plates?
It is possible. The plate is designed to allow separation into individual sections of 8 wells each. The kit includes a sufficient amount of standard for this purpose. However, please be aware that if the plate is used in separate measurements, this necessitates redrawing the calibration curve each time, which, in turn, reduces the total number of samples that can be measured.

For single use → 40 samples (80 wells), with 16 wells used for standards and blank.
For two uses → 32 samples (64 wells), with 32 wells used for standards and blank.

About kit usage

What is the amount of purified water to be added to the standard product?
Because the volume of deionized water (or distilled water) to be added to the freeze-dried standard differs depending on the lot, see the specified volume in a separate sheet. Be sure to allow the solution to stand for 30 minutes, as accurate values may not be obtained.

Overview / Applications

Property

Manufacturer Information

Alias

For research use or further manufacturing use only. Not for use in diagnostic procedures.

Product content may differ from the actual image due to minor specification changes etc.

If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.