It has been added to the cart.

It has been added to the comparison table.

LBIS (TM) Mouse IL-17A ELISA Kit

for Immunochemistry
Manufacturer :
FUJIFILM Wako Pure Chemical Corporation
Storage Condition :
Keep at 2-10 degrees C.
GHS :
  • Structural Formula
  • Label
  • Packing
SDS
Comparison
Product Number
Package Size
Price
Availability
Certificate of Analysis
Purchase
Distributor
293-94701
Barcode No
4548995102942
96Tests
List Price
682.00 USD

In stock in Japan

Document

SDS
Product Specification Sheet
Package Insert
Spectral Data
Certificate of Analysis
Calibration Certificate
Analytical Charts

Kit component

96 tests

Antibody-coated Plate 96 wells/1 plate
Mouse IL-17A Standard 1 bottle
Buffer 60 mL/1 bottle
Biotin-conjugated Antibody 1 bottle
Peroxidase-conjugated Streptavidin Solution 100 μL/1 bottle
TMB Solution 12 mL/1 bottle
Stop Solution 12 mL/1 bottle
Wash Solution (10x) 100 mL/1 bottle
Plate Seal 4 sheets

Product Overview

product img

Interleukin 17A (IL-17A) is an inflammatory cytokine produced primarily by immune cells including Th17 cells. It acts on a diverse range of cells ― including macrophages, fibroblasts, epithelial cells, endothelial cells, and stromal cells ― and efficiently induces the production of inflammatory mediators in these cells.

LBIS™ Mouse IL-17A ELISA Kit is a sandwich ELISA kit for the quantitative measurement of mouse IL-17A. It enables highly sensitive quantification of IL-17A in biological samples. The standard included in the kit is a recombinant protein produced in Escherichia coli.

Kit Performance

Assay target IL-17A
Calibration curve range 2.06 – 800 pg/mL
Analysis sample Mouse Serum/Plasma (EDTA, heparin)
Sample amount 25 μL (2x dilution)
Measurement duration Approx 3 hours and 50 minutes
Wavelength Primary wavelength : 450 nm
Reference wavelength : 620 nm

検量線例

Example of Calibration Curve

Data

Repeatability (within run precision)

Repeatability was conducted by measuring fivefold assays of mouse serum samples at two concentrations.

n\ID ID1 (pg/mL) ID2 (pg/mL)
1 261 18.6
2 248 17.8
3 243 18.4
4 264 18.2
5 274 19.2
mean 258 18.4
SD 12.5 0.52
CV(%) 4.8 2.8
[Result]
The CV (%) of mouse serum was 2.8-4.8%, indicating good repeatability.

Reproducibility (between run precision)

Reproducibility was conducted by measuring mouse serum samples at three concentrations each for four days.

Day\ID ID3 (pg/mL) ID4 (pg/mL) ID5 (pg/mL)
1 281 40.6 5.92
2 309 39.4 5.65
3 308 41.2 6.11
4 297 41.8 5.98
mean 299 40.8 5.92
SD 13.0 1.0 0.19
CV(%) 4.4 2.5 3.3
[Result]
The CV (%) of mouse serum was 2.5-4.4%, indicating good reproducibility.

Dilution Linearity Test

Mouse serum and plasma (EDTA/heparin) were serially diluted twofold using the provided buffer and measured in duplicate to evaluate linearity.

Mouse serum/plasma (EDTA, heparin)

Human serum/plasma (EDTA, heparin)
[Result]
All measurement results showed good linearity within the measurement range.

Spike and Recovery Test

Standard solutions at three concentrations were added to mouse serum/plasma (EDTA/heparin) were conducted, with each measurement performed in duplicate.

Mouse serum

Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery amount
(pg/mL)
Recovery
(%)
- 1.41 - -
10 11.2 9.79 97.9
50 49.1 47.7 95.4
250 264 263 105
mean 99.4

Mouse plasma (EDTA)

Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery amount
(pg/mL)
Recovery
(%)
- 0.328 - -
10 9.99 9.66 96.6
50 50.6 50.3 101
250 257 257 103
mean 100

Mouse plasma (heparin)

Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery amount
(pg/mL)
Recovery
(%)
- 0.749 - -
10 10.9 10.2 102
50 54.6 53.9 108
250 264 263 105
mean 105
[Result]
Good recovery rates were confirmed.

Example of Sample Measurement

Mouse serum and plasma (EDTA/heparin) samples were measured and compared using this kit and competitor Company A.

Mouse serum
(BALB/c, male, 8-week-old, 16-hour fast)

ID\n Fujifilm Wako
(pg/mL)
Company A
(pg/mL)
1 5.75 N.D.
2 46.6 N.D.
3 45.2 5.14
4 1.90 N.D.
5 53.2 N.D.

Mouse plasma (EDTA)
(BALB/c, male, 8-week-old, 16-hour fast)

ID\n Fujifilm Wako
(pg/mL)
Company A
(pg/mL)
1 96.8 N.D.
2 4.28 N.D.
3 12.4 N.D.
4 3.01 N.D.
5 5.62 5.21

Mouse plasma (heparin)
BALB/c, male, 8-week-old, 16-hour fast

ID\n Fujifilm Wako
(pg/mL)
Company A
(pg/mL)
1 57.5 N.D.
2 26.7 N.D.
3 47.2 N.D.
4 6.88 N.D.
5 18.2 N.D.

N.D.: Below the lower limit of quantification
Red: Values outside the quantification range; shown as reference

[Result]
This product was able to detect IL-17A even in samples that were below the lower limit of quantification of another manufacturer's kit.

FAQ

About samples

What anticoagulants should I use?
We recommend using EDTA or heparin as anticoagulant.
How should I store and prepare the samples?
Samples should be immediately assayed or stored below -35℃ until assay. Before starting assay, shake thawed samples sufficiently. Do not repeat freeze-and-thaw cycles. Dilution of a sample should be made in a PP or PE test tube using the provided buffer prior to adding them to wells.

About kit usage

What instruments, and equipment are required for the assay using this kit?
The instruments and equipment required for the use of this kit are listed below;
・Deionized water (or distilled water)
・Test tubes for preparation of standard solution series
・Glassware for dilution of Wash Solution (10x) (a graduated cylinder, a bottle)
・Pipettes (disposable tip type). One should be able to deliver 50 μL precisely, and another for 100 – 1,000 μL
・Syringe-type repeating dispenser like Eppendorf ultipette plus which can dispense 100 μL
・Paper towel to remove washing buffer remaining in wells
・A vortex-type mixer
・A shaker for 96 well-plate (600 – 1,200 rpm)
・An automatic washer for 96 well-plate (if available), or a wash bottle with a jet nozzle
・A 96 well-plate reader (450 ± 10 nm, 620 nm :600 - 650 nm)
・Software for data analysis
How should I use the mouse IL-17A standard?
The amount of purified water to be added varies by lot. Please reconstitute the mouse IL-17A standard with the amount of purified water specified in the “Reconstitution of standard” section on this product page to prepare the original standard solution (8 ng/mL). Then, dilute the solution using the kit-provided buffer that has been equilibrated to room temperature.
What are the storage conditions for each reagent?
Buffer & TMB Solution
Use only volume you need for your assay. Remaining reagents should be stored at 2-10℃ closing the cap tightly. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.
Stop Solution
Close the cap tightly and store at 2-10℃.
Mouse IL-17A Standard & Biotin-conjugated Antibody
The reconstituted standard solution (original standard solution) should be used within 2 weeks if stored in a refrigerator. Dispose the remaining diluted standard solutions after use.
Peroxidase-conjugated Streptavidin Solution
Remaining working solution (already diluted) should be disposed. The rest of the undiluted solution (unused): close the cap tightly and store at 2-10℃. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.

Prepare reagent solutions in appropriate volume for your assay.

Overview / Applications

Property

Manufacturer Information

Alias

  • 637-44711 (Previous Code)
    AKMIL17-011 (Previous Code)

For research use or further manufacturing use only. Not for use in diagnostic procedures.

Product content may differ from the actual image due to minor specification changes etc.

If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.