LBIS (TM) Mouse IFN-γ ELISA Kit
- for Immunochemistry
- Manufacturer :
- FUJIFILM Wako Pure Chemical Corporation
- Storage Condition :
- Keep at 2-10 degrees C.
- GHS :
-
- Structural Formula
- Label
- Packing
- SDS
|
Comparison
|
Product Number
|
Package Size
|
Price
|
Availability
|
Certificate of Analysis
|
Purchase |
|---|---|---|---|---|---|---|
|
|
|
96Tests
|
|
In stock in Japan |
※Check availability in the US with the distributor.
Document
Kit component
96 tests
| Antibody-coated Plate | 96 wells/1 plate |
|---|---|
| Mouse IFN-γ Standard | 1 bottle |
| Buffer | 60 mL/1 bottle |
| Biotin-conjugated Antibody | 1 bottle |
| Peroxidase-conjugated Streptavidin Solution | 100 μL/1 bottle |
| TMB Solution | 12 mL/1 bottle |
| Stop Solution | 12 mL/1 bottle |
| Wash Solution (10x) | 100 mL/1 bottle |
| Plate Seal | 4 sheets |
Product Overview

Interferon-γ (IFN-γ) is an important cytokine involved in both innate and adaptive immunity. Cells known to produce IFN-γ include natural killer (NK) cells and T cells. Secreted IFN-γ induces a wide range of immune responses, including antiviral activity, activation of macrophages, maturation of dendritic cells, and induction of T-cell differentiation. IFN-γ also plays a significant role in antitumor immunity, with reported effects including suppression of metastasis, induction of apoptosis, and inhibition of angiogenesis.
LBIS™ Mouse IFN-γ ELISA Kit is a sandwich ELISA kit for the quantitative measurement of mouse IFN-γ. This kit enables highly sensitive quantification of IFN-γ in biological samples. The standard included in the kit is a recombinant protein produced in Escherichia coli.
Kit Performance
| Assay target | IFN-γ |
|---|---|
| Calibration curve range | 2.05 – 500 pg/mL |
| Analysis sample | Mouse Serum/Plasma (EDTA, heparin) |
| Sample amount | 25 μL (2x dilution) |
| Measurement duration | Approx 3 hours and 50 minutes |
| Wavelength | Primary wavelength : 450 nm Reference wavelength : 620 nm |
Example of Calibration Curve

Data
Repeatability (within run precision)
Repeatability was conducted by measuring fivefold assays of mouse serum samples at two concentrations.
| n\ID | ID1 (pg/mL) | ID2 (pg/mL) |
|---|---|---|
| 1 | 288 | 31.6 |
| 2 | 281 | 30.8 |
| 3 | 276 | 30.2 |
| 4 | 281 | 31.2 |
| 5 | 275 | 29.6 |
| mean | 280 | 30.7 |
| SD | 5.2 | 0.79 |
| CV(%) | 1.8 | 2.6 |
- [Result]
- The CV (%) of mouse serum was 1.8-2.6%, indicating good repeatability.
Reproducibility (between run precision)
Reproducibility was conducted by measuring mouse serum samples at three concentrations each for four days.
| Day\ID | ID3 (pg/mL) | ID4 (pg/mL) | ID5 (pg/mL) |
|---|---|---|---|
| 1 | 200 | 32.0 | 5.22 |
| 2 | 193 | 32.2 | 5.16 |
| 3 | 192 | 33.0 | 4.91 |
| 4 | 195 | 32.7 | 4.89 |
| mean | 195 | 32.5 | 5.05 |
| SD | 3.6 | 0.46 | 0.17 |
| CV(%) | 1.8 | 1.4 | 3.4 |
- [Result]
- The CV (%) of mouse serum was 1.4-3.4%, indicating good reproducibility.
Dilution Linearity Test
Mouse serum and plasma (EDTA/heparin) were serially diluted twofold using the provided buffer and measured in duplicate to evaluate linearity.
Mouse serum

Mouse plasma (EDTA)

Mouse plasma (heparin)

- [Result]
- All measurement results showed good linearity within the measurement range.
Spike and Recovery Test
Standard solutions at three concentrations were added to mouse serum and plasma (EDTA/heparin), then spike recovery tests were conducted, with each measurement performed in duplicate.
Mouse serum
| Spiked amount (pg/mL) |
Measurement value (pg/mL) |
Recovery amount (pg/mL) |
Recovery (%) |
|---|---|---|---|
| - | 8.70 | - | - |
| 20.0 | 27.9 | 19.2 | 96.0 |
| 80.0 | 95.5 | 86.8 | 109 |
| 165 | 174 | 165 | 100 |
| mean | 102 | ||
Mouse plasma (EDTA)
| Spiked amount (pg/mL) |
Measurement value (pg/mL) |
Recovery amount (pg/mL) |
Recovery (%) |
|---|---|---|---|
| - | 4.14 | - | - |
| 20.0 | 24.3 | 20.2 | 101 |
| 80.0 | 84.3 | 80.2 | 100 |
| 165 | 173 | 169 | 102 |
| mean | 101 | ||
Mouse plasma (heparin)
| Spiked amount (pg/mL) |
Measurement value (pg/mL) |
Recovery amount (pg/mL) |
Recovery (%) |
|---|---|---|---|
| - | 4.00 | - | - |
| 20.0 | 24.5 | 20.5 | 103 |
| 80.0 | 87.7 | 83.7 | 105 |
| 165 | 169 | 165 | 100 |
| mean | 103 | ||
- [Result]
- Good recovery rates were confirmed.
Example of Sample Measurement
Mouse serum and plasma (EDTA/heparin) samples were measured and compared using this kit and competitor Company A.
Mouse serum
| ID\n | Fujifilm Wako(pg/mL) | Company A(pg/mL) |
|---|---|---|
| 11 | 30.1 | N.D. |
| 12 | 31.3 | N.D. |
| 13 | 5.73 | N.D. |
| 14 | 56.0 | N.D. |
| 15 | 19.3 | N.D. |
| mean | 28.5 | - |
Mouse plasma (EDTA)
| ID\n | Fujifilm Wako(pg/mL) | Company A(pg/mL) |
|---|---|---|
| 11 | 1.85 | N.D. |
| 12 | 21.7 | N.D. |
| 13 | 38.0 | N.D. |
| 14 | 57.3 | N.D. |
| 15 | 5.85 | N.D. |
| mean | 30.7 | - |
Mouse plasma (heparin)
| ID\n | Fujifilm Wako(pg/mL) | Company A(pg/mL) |
|---|---|---|
| 11 | 61.2 | N.D. |
| 12 | 18.3 | N.D. |
| 13 | 2.66 | 3.09 |
| 14 | 33.1 | N.D. |
| 15 | 77.6 | 4.46 |
| mean | 38.6 | - |
N.D.: Below the lower limit of quantification
Red: Values outside the quantification range; shown as reference
- [Result]
- This product was able to detect IFN-γ even in samples that were below the lower limit of quantification of another manufacturer's kit.
FAQ
About samples
- What anticoagulants should I use?
- We recommend using EDTA or heparin as anticoagulant.
- How should I store and prepare the samples?
- Samples should be immediately assayed or stored below -35℃ until assay. Before starting assay, shake thawed samples sufficiently. Do not repeat freeze-and-thaw cycles. Dilution of a sample should be made in a PP or PE test tube using the provided buffer prior to adding them to wells.
About kit usage
- What instruments, and equipment are required for the assay using this kit?
- The instruments and equipment required for the use of this kit are listed below;
・Deionized water (or distilled water)
・Test tubes for preparation of standard solution series
・Glassware for dilution of Wash Solution (10x) (a graduated cylinder, a bottle)
・Pipettes (disposable tip type). One should be able to deliver 50 μL precisely, and another for 100 – 1,000 μL
・Syringe-type repeating dispenser like Eppendorf multipette plus which can dispense 100 μL
・Paper towel to remove washing buffer remaining in wells
・A vortex-type mixer
・A shaker for 96 well-plate (600 – 1,200 rpm)
・An automatic washer for 96 well-plate (if available), or a wash bottle with a jet nozzle
・A 96 well-plate reader (450 ± 10 nm, 620 nm :600 - 650 nm)
・Software for data analysis
- How should I use the mouse IFN-γ standard?
- The amount of purified water to be added varies by lot. Please reconstitute the mouse IFN-γ standard with the amount of purified water specified in the “Reconstitution of standard” section on this product page to prepare the original standard solution (5 ng/mL). Then, dilute the solution using the kit-provided buffer that has been equilibrated to room temperature.
- What are the storage conditions for each reagent?
- Buffer & TMB Solution
Use only volume you need for your assay. Remaining reagents should be stored at 2-10℃ closing the cap tightly. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.
Stop Solution
Close the cap tightly and store at 2-10℃.
Mouse IFN-γ Standard & Biotin-conjugated Antibody
The reconstituted standard solution (original standard solution) should be used within 2 weeks if stored in a refrigerator. Dispose the remaining diluted standard solutions after use.
Peroxidase-conjugated Streptavidin Solution
Remaining working solution (already diluted) should be disposed. The rest of the undiluted solution (unused): close the cap tightly and store at 2-10℃. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.
Prepare reagent solutions in appropriate volume for your assay.
Overview / Applications
Property
Manufacturer Information
Alias
- 630-44701 (Previous code)
AKMIFNG-011 (Previous code)
For research use or further manufacturing use only. Not for use in diagnostic procedures.
Product content may differ from the actual image due to minor specification changes etc.
If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.