It has been added to the cart.

It has been added to the comparison table.

LBIS™ Mouse MCP-1 (CCL2) ELISA Kit

for Immunochemistry
Manufacturer :
FUJIFILM Wako Pure Chemical Corporation
Storage Condition :
Keep at 2-10 degrees C.
GHS :
  • Structural Formula
  • Label
  • Packing
SDS
Comparison
Product Number
Package Size
Price
Availability
Certificate of Analysis
Purchase
Distributor
290-88101
Barcode No
4548995102461
96Tests
List Price
682.00 USD

In stock in Japan

Document

SDS
Product Specification Sheet
Package Insert
Spectral Data
Certificate of Analysis
Calibration Certificate
Analytical Charts

Kit component

96 tests

Antibody-coated Plate 96 wells/1 plate
Mouse MCP-1 (CCL2) Standard 1 bottle
Buffer 60 mL/1 bottle
Biotin-conjugated Antibody 1 bottle
Peroxidase-conjugated Streptavidin Solution 100 μL/1 bottle
TMB Solution 12 mL/1 bottle
Stop Solution 12 mL/1 bottle
Wash Solution (10x) 100 mL/1 bottle
Plate Seal 4 sheets

Product Overview

product img

MCP-1 (CCL2) is a member of the CC chemokine family and exhibits chemotactic activity toward monocytes and macrophages. Because MCP-1 and its receptor recruit these cells to sites of inflammation--thereby sustaining inflammatory responses--they are implicated in metabolic syndrome and cardiovascular diseases.
LBIS™ Mouse MCP-1 (CCL2) ELISA Kit is a sandwich ELISA kit designed for the quantitative measurement of mouse MCP-1 (CCL2). This kit enables highly sensitive quantification of MCP-1 (CCL2) in biological samples. The standard included in the kit is a recombinant protein produced in Escherichia coli.

Kit Performance

Assay target MCP-1 (CCL2)
Calibration curve range 3.85 – 500 pg/mL
Analysis sample Mouse Serum/Plasma (EDTA/heparin)/Urine/ Culture medium (D-MEM/RPMI-1640 etc.)*
Sample amount 10 μL (5x dilution)
Measurement duration Approx 3 hours and 50 minutes
Wavelength Primary wavelength : 450 nm
Reference wavelength : 620 nm
*It is tested by using the samples that standard substances have been added to the culture medium.
Measurement availability depends on the culture medium.
When using cell culture supernatant as a sample, please make sure to perform recovery test and dilution linearity test on the culture medium before use.

Example of Calibration Curve

Example of Calibration Curve

Data

Repeatability (within run precision)

Repeatability was conducted by measuring fivefold assays of BALB/c mouse serum samples at two concentrations.

n\ID ID1 (pg/mL) ID2 (pg/mL)
1 170 18.1
2 171 18.9
3 172 18.7
4 173 18.4
5 169 18.8
mean 171 18.6
SD 1.6 0.33
CV(%) 0.92 1.8
[Result]
The CV (%) of mouse serum was 0.92-1.8%, indicating good repeatability.

Reproducibility (between run precision)

Reproducibility was conducted by measuring BALB/c mouse serum samples at three concentrations each for four days.

Day\ID ID3 (pg/mL) ID4 (pg/mL) ID5 (pg/mL)
1 210 50.8 8.67
2 211 48.0 8.33
3 215 48.2 8.25
4 226 51.9 8.84
mean 216 49.7 8.52
SD 7.3 1.9 0.28
CV(%) 3.4 3.9 3.3
[Result]
The CV (%) of mouse serum was 3.3-3.9%, indicating good reproducibility.

Dilution Linearity Test

BALB/c mouse serum and plasma (EDTA/heparin), as well as urine samples, were serially diluted twofold using the provided buffer and measured in duplicate to evaluate linearity.

Mouse serum

Mouse serum

Mouse plasma (EDTA)

Mouse plasma (EDTA)

Mouse plasma (heparin)

Mouse plasma (heparin)

Mouse urine

Mouse urine
[Result]
All measurement results showed good linearity within the measurement range.

Spike and Recovery Test

Standard solutions at three concentrations were added to Balb/c mouse serum/plasma (EDTA/heparin), urine, and culture medium then spike recovery tests were conducted, with each measurement performed in duplicate.

Mouse serum

Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery amount
(pg/mL)
Recovery
(%)
- 25.7 - -
25.0 50.7 25.0 100
50.0 75.5 49.8 99.6
100 124 98.3 98.3
200 223 197 98.5
mean 99.1

Mouse plasma (EDTA)

Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery amount
(pg/mL)
Recovery
(%)
- 14.7 - -
25.0 40.5 25.8 103
50.0 66.5 51.8 104
100 115 100 100
200 205 190 95.0
mean 101

Mouse plasma (heparin)

Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery amount
(pg/mL)
Recovery
(%)
- 17.3 - -
25.0 42.4 25.1 100
50.0 67.8 50.5 101
100 118 101 101
200 217 200 100
mean 101

Culture medium (SFM/CD/GIT/D-MEM/RPMI-1640)

SFM CD GIT
Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery
(%)
Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery
(%)
Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery
(%)
0 0 0
6.25 6.39 102 6.25 7.12 114 6.25 6.38 102
25.0 26.1 104 25.0 25.0 100 25.0 26.9 108
50.0 50.6 101 50.0 46.5 92.9 50.0 50.3 101
100 96.0 96.0 100 87.2 87.2 100 96.4 96.4
mean 101 mean 98.5 mean 102
D-MEM RPMI-1640
Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery
(%)
Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery
(%)
0 0
6.25 6.57 105 6.25 6.44 103
25.0 26.7 107 25.0 29.2 117
50.0 50.5 101 50.0 54.0 108
100 101 101 100 103 103
mean 104 mean 108
[Result]
Good recovery rates were confirmed.

Correlation with conventional product

For BALB/c mouse serum and plasma (EDTA/heparin) samples, measurements were conducted using this kit and competitor Company A. The measurement values obtained using both products were then compared.

Sample measurement value

Sample measurement value

Correlation

Correlation
[Result]
Good correlation was confirmed.

FAQ

About samples

What anticoagulants should I use?
We recommend using EDTA or heparin as anticoagulant.
How should I store and prepare the samples?
Samples should be immediately assayed or stored below -35℃ until assay. Before starting assay, shake thawed samples sufficiently. Do not repeat freeze-and-thaw cycles.
Dilution of a sample should be made in a PP or PE test tube using the provided buffer prior to adding them to wells.

About kit usage

What instruments, and equipment are required for the assay using this kit?
The instruments and equipment required for the use of this kit are listed below;
・Deionized water (or distilled water)
・Test tubes for preparation of standard solution series
・Glassware for dilution of Wash Solution (10x) (a graduated cylinder, a bottle)
・Pipettes (disposable tip type). One should be able to deliver 50 μL precisely, and another for 100 – 1,000 μL
・Syringe-type repeating dispenser like Eppendorf multipette plus which can dispense 100 μL
・Paper towel to remove washing buffer remaining in wells
・A vortex-type mixer
・A shaker for 96 well-plate (600 – 1,200 rpm)
・An automatic washer for 96 well-plate (if available), or a wash bottle with a jet nozzle
・A 96 well-plate reader (450 ± 10 nm, 620 nm :600 - 650 nm)
・Software for data analysis
How should I use the mouse MCP-1 standard?
The amount of purified water to be added varies by lot. Please reconstitute the mouse MCP-1 standard with the amount of purified water specified in the “Reconstitution of standard” section on this product page to prepare the original standard solution (5,000 pg/mL). Then, dilute the solution using the kit-provided buffer that has been equilibrated to room temperature.
What are the storage conditions for each reagent?
Buffer & TMB Solution
Use only volume you need for your assay. Remaining reagents should be stored at 2-10℃ closing the cap tightly. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.
Stop Solution
Close the cap tightly and store at 2-10℃.
Mouse MCP-1 Standard & Biotin-conjugated Antibody
Mouse MCP-1 Standard & Biotin-conjugated Antibody The reconstituted standard solution (original standard solution) should be used within 2 weeks if stored in a refrigerator. Dispose the remaining diluted standard solutions after use.
Peroxidase-conjugated Streptavidin Solution
Remaining working solution (already diluted) should be disposed. The rest of the undiluted solution (unused): close the cap tightly and store at 2-10℃. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.

Remaining working solution (already diluted) should be disposed. The rest of the undiluted solution (unused): close the cap tightly and store at 2-10℃. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.

Overview / Applications

Property

Manufacturer Information

Alias

  • 637-54101 (Previous code)
    AKMMCP1-011 (Previous code)

For research use or further manufacturing use only. Not for use in diagnostic procedures.

Product content may differ from the actual image due to minor specification changes etc.

If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.