Product Information
Secretory IgA (s-IgA) is an IgA to which a secretory component (SC) is bound, basically occurring as a dimer but also as tetramer, etc. IgA in secretions occurs principally as s-IgA, which plays an important role in protection from infection on the mucosa of the respiratory and gastrointestinal tracts. s-IgA is also found in human saliva and produced in plasma cells in the salivary glands. The level of s-IgA in human saliva is reportedly increased by acute stress1) and decreased by chronic stress2) 3), giving it utility as a stress biomarker.
Human s-IgA ELISA Kit Wako is an ELISA kit for assaying s-IgA in human saliva.
-
Kit Performance
The range of the calibration curve 0.082 - 20 µg/mL Assay target Secretory IgA (s-IgA) Sample Human Saliva The sample amount required 5 µL Assay time 2.5 hours Detection method Colorimetry Within assay variation CV < 6% Between assay variation CV < 12% -
Example of Calibration Curve
Data
-
Spiked recovery test
Sample Amount spiked(µg/mL) Measured value(µg/mL) Recovery rate(%) Saliva1 0 64.8 - 40 108.8 103.8 160 220.8 98.2 400 404.8 87.1 Saliva 2 0 46.4 - 40 96.0 111.1 160 196.0 95.0 400 432.0 96.8 Saliva 3 0 120.8 - 40 189.6 117.9 160 274.4 97.7 400 476.0 91.4 Saliva 4 0 112.0 - 40 157.6 103.7 160 244.8 90.0 400 441.6 86.3 Saliva 5 0 105.6 - 40 152.0 104.4 160 250.4 94.3 400 600.8 118.8 -
Dilution linearity test
Sample Dilution Measured value(µg/mL) Saliva1 1/1 363.5 1/2 201.7 1/4 107.7 1/8 53.6 Saliva 2 1/1 237.4 1/2 123.9 1/4 65.3 1/8 32.5 Saliva 3 1/1 310.0 1/2 167.3 1/4 89.5 1/8 42.7 Saliva 4 1/1 293.3 1/2 145.3 1/4 74.7 1/8 38.3 Saliva 5 1/1 188.2 1/2 97.5 1/4 47.4 1/8 26.8
Cross reactivity with various types of immunoglobulins
| Immunoglobulins | Cross-reactivity(%) |
|---|---|
| Serum IgG(Human) | < 0.1 |
| Serum IgA(Human) | 0.3 |
| Serum IgM(Human) | 0.0 |
| Serum IgE(Human) | 0.0 |
References
- Viena, T. D. et al.: Biol. Psychol., 91(2), 307(2012).
Differential effects of mild chronic stress on cortisol and S-IgA responses to an acute stressor - Deinzer, R. et al.: Int. J. Psychophysiol., 37(3), 219(2000).
Prolonged reduction of salivary immunoglobulin A (sIgA) after a major academic exam - Phillips, A. C. et al.: Brain, behav. Immun., 20(2), 191(2006).
Stressful life events are associated with low secretion rates of immunoglobulin A in saliva in the middle aged and elderly
FAQ
About sample
- How should I store samples?
- Saliva samples should be stored frozen at -30°C or below. Avoid repeated freezing and thawing of samples.
- How should I prepare samples?
- Dilute the saliva sample 80-fold with a buffer previously diluted with distilled water (395 μL of diluted buffer : 5 μL of saliva sample). If the saliva sample is stored frozen, thaw at room temperature and stir well. If the concentration of the diluted sample is higher than 20 μg/mL, further dilute the 80-fold diluted sample with the buffer before measurement.
- How should I pretreat highly viscous saliva samples?
- Highly viscous saliva may not provide a correct measurement value. Precipitate mucin, etc. by centrifugation and use the supernatant.
About analyte
- What is the degree of cross-reactivity with non-human species?
- Fujifilm Wako has not checked cross-reactivity with non-human species (mice, rats, etc.).
About the kit
- What is the origin of the standard?
- The origin is a purified product from human colostrum.
- Can I divide plates?
- Since the plate in this kit are separable and each row (8 wells) is detachable, divided use of the kit is possible.
About kit usage
- What reagents, instruments, and equipment are required for measurement with this kit?
- The following reagents, instruments, and equipment are required to use this kit:
- Micropipettes and disposable tips (5 to 1,000μL) (8- or 12- channel pipettes are recommended.)
- Microplate reader which can read extinction 3.0 at 450 nm
- Microplate shaker or conventional shaker
- Test tubes for dilution of standard solutions and samples (glass or polypropylene)
- Microplate washer (For manual operation, a continuous dispenser, needle dispenser, and an aspirator or vacuum pump are recommended.)
- Graduated cylinder (1,000 mL)
- Distilled or deionized water
Overview / Applications
Property
Manufacturer Information
Alias
For research use or further manufacturing use only. Not for use in diagnostic procedures.
Product content may differ from the actual image due to minor specification changes etc.
If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.