PS Capture(TM) Exosome ELISA Kit (Anti Mouse IgG POD)
- for Genetic Research
- Manufacturer :
- FUJIFILM Wako Pure Chemical Corporation
- Storage Condition :
- Keep at 2-10 degrees C.
- GHS :
-
- Structural Formula
- Label
- Packing
Comparison
|
Product Number
|
Package Size
|
Price
|
Availability
|
Certificate of Analysis
|
Purchase |
---|---|---|---|---|---|---|
|
|
96Tests
|
|
Out of Stock |
※Check availability in the US with the distributor.
Document
Kit component
FAQ
Kit specifications and performance
- Do I have to prepare a standard for every assay? And do I have to use Wako MagCapture™ Exosome Isolation Kit PS for preparation of the standard?
- When you perform a quantitative assay, prepare an extracellular vesicle sample as standard. Although an extracellular vesicle sample purified by ultracentrifugation or polymer precipitation may also be used as standard, an extracellular vesicle sample purified by the PS affinity method based on the principle identical with that of the assay is recommended (see the instruction manual included in the kit for details of the preparation method).
- Why does this kit include no standard?
- The standard and assay samples must be derived from identical cell species, because the type and amounts of surface marker proteins on extracellular vesicles secreted may vary depending on the cell type. Therefore, this kit does not include a standard. Prepare a standard purified from culture supernatant of cells identical with the source cells of assay samples.
- Is this kit compatible with direct assay of serum and plasma samples?
- This kit is not recommended for direct assay of serum and plasma samples from human, mouse, and rat, because the secondary antibody for detection included in it reacts with human, mouse, and rat IgG non-specifically. However, this kit is compatible with qualitative assay of extracellular vesicle samples purified from serum and plasma specimens using MagCapture™ Exosome Isolation Kit PS. It is also compatible with qualitative assay of extracellular vesicles purified by ultracentrifugation or polymer precipitation.
- Is this kit compatible with direct assay of cell culture supernatant samples?
- This kit is compatible with direct assay of cell culture supernatant from both serum-free and FBS-supplemented media because primary antibody (Anti- CD63 antibody) and secondary antibody included in the kit don’t react with bovine IgG non-specifically. Utilize this kit for both quantitative and qualitative analyses of extracellular vesicles in cell culture supernatant samples.
- Can I replace the primary antibody with another antibody?
- Yes. Choose a mouse antibody against the surface marker you want to detect and investigate the optimal concentration according to the instruction manual.
- How can I store the remaining reagents?
- See [6. The storage method of each reagent when the kit is separately used] in the instruction manual included in the kit.
Comparison with conventional methods
- Is the detection sensitivity of this kit higher than other methods?
- This kit has been confirmed to detect extracellular vesicles at a sensitivity higher than those of ELISA methods involving antibody immobilization or direct immobilization of purified samples on the plate. In addition, a correlation between ELISA results obtained with this kit and Western Blotting results has been confirmed.
Operational procedures and composition of the kit
- How long is the operation time for ELISA using this kit?
- The entire kit process takes approximately 5 hours, including immobilization of diluted cell culture supernatant specimens or purified and diluted extracellular vesicle samples onto a 96-well plate for 2 hours, reaction with the primary antibody for 1 hour, reaction with the secondary antibody for 1 hour, and reaction with tetramethylbenzidine (TMB) for 30 minutes. With washing and other operations included, the assay is completed in approximately 5 hours. After addition of Stop Solution, measure the absorbance at the main wavelength 450 nm and the complementary wavelength 620 nm (600 - 650 nm).
- Is an Exosome Capture 96 Well Plate compatible with recycling?
- No. An Exosome Capture 96 Well Plate is not compatible with recycling by regeneration, because Stop Solution denatures proteins on the plate.
- Is there any step that can be carried over to next day?
- Immobilization of individual samples onto the plate may be prolonged up to overnight at 4°C without any problem.
Sample amount/volume
- How much is the minimum sample amount required for detection using this kit?
- Extracellular vesicles corresponding to 1 ng protein are detectable using this kit. The detection limit of extracellular vesicles purified from COLO201 cell culture supernatant was 11 pg (it has been confirmed that the detection limit varies depending on the cell strain).
- How much sample volume is required for direct assay of culture supernatant?
- Culture supernatant of a few µL in volume (1-5 µL) is sufficient for assay. This is recommended for monitoring of change in number of extracellular vesicles in culture medium over time and assay of new cell culture supernatant. However, the number of extracellular vesicles in culture medium may be limited depending on the cell species (e.g., iPS cells) and the sample volume per well used for ELISA should be investigated as necessary. If the number of extracellular vesicles in the sample is unknown, use of undiluted culture supernatant samples (100 µL) is recommended.
Related products
- Are there any primary antibodies recommended for detection?
- The following antibodies have been confirmed to be compatible with ELISA at our Laboratories.
Antigen Reactivity Antibody Manufacturer Use CD63 human Mouse anti-CD63 monoclonal antibody (3-13) Wako, Code:012-27063 WB、ELISA CD81 human Mouse anti-CD81 monoclonal antibody (M38) Novus, Code:NBP1-44861
(Wako 550-30161)ELISA CD9 Mouse Rat anti-CD9 monoclonal antibody (MZ3) Bio Legend, Code:124802 ELISA CD63 Mouse Rat anti-CD63 monoclonal antibody (NVG-2) Bio Legend, Code:143902 ELISA CD81 Mouse Armenian hamster anti-CD81 monoclonal antibody (Eat-2) Bio Legend, Code:104902 ELISA
Trouble shooting
- My detection results are not satisfactory. Could you tell me what to check for?
- Check if any of the reagents has been expired. When you fail to detect a positive signal even with Control Primary Antibody Anti CD63 (100 × ) (included in this kit), it may be ascribable to an expression level of CD63 under the detection limit or some other cause. Please inquire to us in such a case.
Overview / Applications
Outline | This product is for research use only. Do not administer it to human. This kit is an enzyme immunoassay reagent that can be used for the qualitative analysis of extracellular vesicles purified from cell culture supernatants and body fluid specimens and quantitative analysis of extracellular vesicles in cell culture supernatant specimens. Extracellular vesicles with optional marker protein on the surface can be detected with high sensitivity by using a solid phase immobilized protein that binds specifically to phosphatidylserine (PS) of extracellular vesicles and an arbitrary detection mouse monoclonal primary antibody. Features 1. High sensitivity detection of extracellular vesicles by Tim 4 affinity method for phosphatidylserine 2. Quantitative measurement of extracellular vesicles in culture supernatant (requires extracellular vesicle standard purified with MagCapture (TM) Exosome Isolation Kit) 3. Sensitivity of 50 to 1000 times or more of the western blot method 4. Detection control Anti-CD63 antibody attached as primary antibody |
---|
Property
Manufacturer Information
Alias
For research use or further manufacturing use only. Not for use in diagnostic procedures.
Product content may differ from the actual image due to minor specification changes etc.
If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.