GLP-1 ELISA Kit Wako, High Sensitive

for Diabetes Research
Manufacturer :
FUJIFILM Wako Pure Chemical Corporation
Storage Condition :
Keep at 2-10 degrees C.
GHS :
  • Structural Formula
  • Label
  • Packing
SDS
Comparison
Product Number
Package Size
Price
Availability
Certificate of Analysis
Purchase
Distributor
299-75501
Barcode No
4548995038357
96Tests
List Price
862.00 USD

Out of Stock

Document

SDS
Product Specification Sheet
Package Insert
Spectral Data
Certificate of Analysis
Calibration Certificate
Analytical Charts

Kit component

For 96 tests

Antibody-coated Plate 96wells (8×12) / 1 plate
GLP-1 Standard 1 bottle
Buffer Solution 60 mL / 1 bottle
Biotinylated Anti GLP-1 Antibody Solution 100 μL / 1 vial
Peroxidase-conjugated Streptavidin Solution 100 μL / 1 vial
TMB Solution 12 mL / 1 bottle
Stop Solution 12 mL / 1 bottle
Wash Solution (×10) 100 mL / 1 bottle
Plate Seal 4 sheets

Product Overview

GLP-1 is a gastrointestinal hormone secreted from intestinal L-cells that lowers blood glucose levels by enhancing insulin secretion and suppressing glucagon secretion in response to glucose stimulation. Reports have indicated that GLP-1 is useful for the treatment of type 2 diabetes as a novel antidiabetic agent.
This product is an ELISA kit that enables highly sensitive measurement of GLP-1 in mouse and rat plasma using a GLP-1 antibody.

Kit Performance

Analysis sample Mouse Plasma (EDTA)
Rat Plasma (EDTA)
Calibration curve range 0.94-30 pM
Sample volume 10 μL
Measurement duration Approx. 4 hours
Detection method Colorimetric
(Primary: 450 nm / Reference: 620 nm)

Example of Calibration Curve

Example of Calibration Curve

Data

Repeatability (within run precision)

Repeatability was conducted by measuring five fold assays of plasma (EDTA) samples from mouse and at two concentrations with this product.

n\ID Mouse (C57BL/6J, 12 weeks old, ♂) plasma (EDTA)
ID1 (pM) ID2 (pM)
1 33.9 7.61
2 34.2 8.01
3 35.1 8.12
4 33.4 8.09
5 33.2 7.95
mean 34.0 7.96
SD 0.750 0.205
CV(%) 2.2 2.6

[Result]
The CV (%) of mouse plasma was 2.2-2.6%, indicating good repeatability.

Reproducibility (between run precision)

Reproducibility was conducted by measuring plasma (EDTA) samples from mouse at three concentrations each for four days with this product.

Day\ID Mouse (C57BL/6J, 12 weeks old, ♂) plasma (EDTA)
ID3 (pM) ID4 (pM) ID5 (pM)
1 70.5 34.0 7.96
2 69.8 32.7 7.55
3 68.4 31.7 7.32
4 71.2 33.9 7.76
mean 70.0 33.1 7.65
SD 1.20 1.09 0.275
CV(%) 1.7 3.3 3.6

[Result]
The CV (%) of mouse plasma was 1.7-3.6%, indicating good reproducibility.

Dilution Linearity Test

Plasma (EDTA) samples from mouse were serially diluted twofold with buffer in this kit and measured in duplicate to assess linearity.

Dilution Linearity Test

[Result]
All measurement results showed good linearity within the measurement range.

Cross-Reactivity with Similar Peptides

Using this product, cross-reactivity was assessed by measuring structurally related proteins at each concentrations.

Species Similar peptides Concentrations
(pg/mL)
Measurement value
(pg/mL)
Cross-Reactivity
(%)
Mouse/ Rat GLP-2 1,000 N.D. -
Glucagon (1-29) 1,000 N.D. -
Insulin 1,000 N.D. -
Secretin 1,000 N.D. -
GIP 1,000 N.D. -
VIP 1,000 N.D. -
GRF 1,000 N.D. -
Bovine Glucagon (1-29) 1,000 N.D. -
VIP 1,000 N.D. -
Porcine Glucagon (1-29) 1,000 N.D. -
VIP 1,000 N.D. -

Correlation with products from other competitors

Measurements were conducted on the same plasma samples (n=10) from mouse using this kit and competitor's product M. The measurement values obtained using both products were then compared.

Correlation with products from other competitors

GLP-1 ELISA Kit Wako, High Sensitive (This kit)

Sample ID Measurement value
(pM)
mean SD CV
(%)
1 30.7 31.4 0.976 3.1
32.1
2 16.0 16.3 0.474 2.9
16.6
3 16.8 17.2 0.640 3.7
17.7
4 21.3 20.8 0.651 3.1
20.4
5 27.2 27.1 0.0884 0.3
27.1
6 40.3 39.6 1.02 2.6
38.9
7 15.5 15.9 0.693 4.3
16.4
8 9.18 9.43 0.354 3.8
9.68
9 7.84 8.09 0.357 4.4
8.35
10 14.1 13.9 0.247 1.8
13.7

Competitor’s product M

Sample ID Measurement value
(pM)
mean SD CV
(%)
1 24.4 23.4 1.38 5.9
22.4
2 11.9 11.4 0.665 5.8
10.9
3 13.5 14.0 0.786 5.6
14.6
4 17.2 16.9 0.448 2.7
16.5
5 20.7 20.5 0.284 1.4
20.3
6 20.1 19.5 0.719 3.7
19.0
7 11.6 11.2 0.668 6.0
10.7
8 7.28 8.87 2.25 25.4
10.5
9 9.50 8.64 1.22 14.1
7.78
10 8.02 8.27 0.350 4.2
8.52

[Result]
Although the results obtained with this kit showed a strong correlation with those obtained with Competitor Product M, discrepancies in the measurement values were observed at high concentrations. Furthermore, the CV for this kit was below 5%, whereas a higher degree of variability was observed with Competitor Product M.

FAQ

About samples

Is there any influence from substances that interfere with measurement?
Although there is no evaluation data within the company, if presence of an interfering substance is suspected, examine with a dilution test at more than 2 points. Before starting the assay, dilute samples 5x in test tubes (PP or PE). Multiply the assay value by this dilution rate (5x) in calculation.

About kit usage

What instruments, and equipment are required for the assay using this kit?
  • Distilled water or deionized water
  • Test tubes for preparation of standard solution series.
  • Glassware for dilution of Wash Solution (a graduated cylinder and a bottle)
  • Pipettes (disposable tip type) One should be able to deliver 50 µL precisely, and another for 100-500 µL.
  • Syringe-type repeating dispenser like Eppendorf multi-pipette plus which can dispense 50 µL.
  • Paper towel to remove washing buffer remaining in wells.
  • A vortex-type mixer.
  • A shaker for 96 well-plate (600-1,200 rpm).
  • An automatic washer for 96 well-plate (if available), or a washing bottle with a jet nozzle.
  • A 96 well-plate reader (450 nm ± 10 nm, 620 nm (600-650 nm)).
  • Software for data analysis, if available.
What are the precautions for the solution preparation or measurement?
Please read the instruction manual carefully before measurement to confirm sample conditions, measurement conditions, and measurement methods.
Allow the plate and reagents to equilibrate to room temperature (20°C - 25°C) for approximately 2 hours.
Dilute the wash solution 10-fold with purified water that has been brought to room temperature.
Dilute the standard solution with buffer in this kit that has been brought to room temperature.
What are the storage conditions guidelines for each reagent?

Antibody-coated Plate
If seal is not removed, put the strip back in a plastic bag with zip-seal originally used for well-plate container and store at 2°C - 10°C.

GLP-1 Standard
Do not allow the undiluted solution to return to room temperature. Immediately fasten the cap tightly and store at 2°C - 10°C. Please use within two weeks. Standard solutions prepared above should be used as soon as possible, and should not be stored. Dispose remaining prepared solution.

Buffer & TMB Solution
Use only volume you need for your assay. Remaining reagents should be stored at 2°C - 10°C fastening the cap tightly. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.

Biotinylated Anti GLP-1 Antibody Solution & Peroxidase-conjugated Streptavidin Solution
Unused working solution (already diluted) should be disposed. Undiluted solution should be fastened the cap and stored at 2°C - 10°C.

Stop Solution
Remaining solution should be fastened the cap and stored at 2°C - 10°C.

Wash Solution (x10)
Undiluted solution should be fastened the cap and stored at 2°C - 10°C. Dispose any unused diluted wash solution.

Overview / Applications

Outline GLP-1 is a gastrointestinal hormone secreted from intestinal L-cells that lowers blood glucose levels by enhancing insulin secretion and suppressing glucagon secretion in response to glucose stimulation. Reports have indicated that GLP-1 is useful for the treatment of type 2 diabetes as a novel antidiabetic agent.
This product is an ELISA kit that enables highly sensitive measurement of GLP-1 in mouse and rat plasma using a GLP-1 antibody.

Property

Manufacturer Information

Alias

For research use or further manufacturing use only. Not for use in diagnostic procedures.

Product content may differ from the actual image due to minor specification changes etc.

If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.