LBIS (TM) Human IL-7 ELISA Kit

for Immunochemistry
Manufacturer :
FUJIFILM Wako Pure Chemical Corporation
Storage Condition :
Keep at 2-10 degrees C.
  • Structural Formula
  • Label
  • Packing
Comparison
Product Number
Package Size
Price
Inventory
Distributor
290-87501
Barcode No
4548995102935
96Tests
Inquire

Document

Product Specification Sheet
Package Insert
Spectral Data
Certificate of Analysis
Calibration Certificate
Analytical Charts

Kit component

96 tests

Antibody-coated Plate 96 wells/1 plate
Human IL-7 Standard 1 bottle
Buffer 60 mL/1 bottle
Biotin-conjugated Antibody 1 bottle
Peroxidase-conjugated Streptavidin Solution 150 μL/1 bottle
TMB Solution 12 mL/1 bottle
Stop Solution 12 mL/1 bottle
Wash Solution (10x) 100 mL/1 bottle
Plate Seal 4 sheets

Product Overview

Human Interleukin-7 (IL-7) - image

Human Interleukin-7 (IL-7) is a glycoprotein with a molecular weight of approximately 25,000 Da, composed of 152 amino acid residues, and is an essential cytokine for the adaptive immune system. IL-7 promotes the differentiation of immune cells such as natural killer (NK) cells, B cells, and T cells, as well as host defense. It is involved in chronic infectious diseases, rheumatoid arthritis, asthma, and other conditions.

LBIS™ Human IL-7 ELISA Kit is a sandwich ELISA kit for the quantitative measurement of human IL-7. This kit enables highly sensitive quantification of IL-7 in biological samples. The standard included in the kit is a recombinant protein produced in Escherichia coli.

Kit Performance

Assay target IL-7
Calibration curve range 1.50 - 236 pg/mL
Analysis sample Human Serum/Plasma (EDTA, heparin)
Sample amount 50 μL (2x dilution)
Measurement duration Approx 2 hours and 50 minutes
Wavelength Primary wavelength : 450 nm
Reference wavelength : 620 nm

Example of Calibration Curve

Example of Calibration Curve

Data

Repeatability (within run precision)

Repeatability was conducted by measuring fivefold assays of human serum samples at two concentrations.

n\ID ID1 (pg/mL) ID2 (pg/mL)
1 2.00 30.4
2 2.13 31.6
3 2.21 31.5
4 2.14 32.1
5 2.14 31.3
mean 2.12 31.4
SD 0.076 0.62
CV(%) 3.6 2.0
[Result]
The CV (%) of human serum was 2.0-3.6%, indicating good repeatability.

Reproducibility (between run precision)

Reproducibility was conducted by measuring human serum samples at three concentrations each for four days.

Day\ID ID3 (pg/mL) ID4 (pg/mL) ID5 (pg/mL)
1 124 7.72 4.08
2 126 7.93 3.99
3 129 7.86 4.18
4 121 7.64 3.88
mean 125 7.79 4.03
SD 3.4 0.13 0.13
CV(%) 2.7 1.7 3.2
[Result]
The CV (%) of human serum was 1.7-3.2%, indicating good reproducibility.

Dilution Linearity Test

Human serum and plasma (EDTA/heparin) were serially diluted twofold using the provided buffer and measured in duplicate to evaluate linearity.

Human serum

Human serum

Human plasma (EDTA)

Human plasma (EDTA)

Human plasma (heparin)

Human plasma (heparin)
[Result]
All measurement results showed good linearity within the measurement range.

Spike and Recovery Test

Standard solutions at three concentrations were added to human serum/plasma (EDTA), then spike recovery tests were conducted, with each measurement performed in duplicate.

Human serum

Spiked amount (pg/mL) Measurement value (pg/mL) Recovery amount (pg/mL) Recovery (%)
- 8.37 - -
12.5 19.9 11.5 92.0
25.0 31.9 23.5 94.0
50.0 57.5 49.1 98.2
100 113 105 105
mean 97.3

Human plasma (EDTA)

Spiked amount (pg/mL) Measurement value (pg/mL) Recovery amount (pg/mL) Recovery (%)
- 8.33 - -
12.5 19.9 11.6 92.8
25.0 32.4 24.1 96.4
50.0 57.7 49.4 98.8
100 109 101 101
mean 96.7

Human plasma (heparin)

Spiked amount (pg/mL) Measurement value (pg/mL) Recovery amount (pg/mL) Recovery (%)
- 8.65 - -
12.5 21.2 12.6 101
25.0 33.6 25.0 100
50.0 58.9 50.3 101
100 113 104 104
mean 101
[Result]
Good recovery rates were confirmed.

Example of Sample Measurement

Human serum samples were measured and compared using this kit.

ID\n Measurement value (pg/mL) mean SD CV (%)
1 2
1 3.85 3.95 3.90 0.071 1.8
2 2.53 2.36 2.45 0.12 4.9
3 7.83 7.35 7.59 0.34 4.5
4 6.39 6.03 6.21 0.25 4.1
5 9.55 9.18 9.37 0.26 2.8
6 5.53 5.17 5.35 0.25 4.8
7 3.69 3.86 3.78 0.12 3.2
8 10.7 11.3 11.0 0.42 3.9
9 39.2 40.8 40.0 1.1 2.8
10 9.85 10.2 10.0 0.25 2.5
11 3.41 3.58 3.50 0.12 3.4
[Result]
This product was able to detect IL-7.

FAQ

About samples

What anticoagulants should I use?
We recommend using EDTA or heparin as anticoagulant.
How should I store and prepare the samples?
Samples should be immediately assayed or stored below -35℃ until assay. Before starting assay, shake thawed samples sufficiently. Do not repeat freeze-and-thaw cycles. Dilution of a sample should be made in a PP or PE test tube using the provided buffer prior to adding them to wells.

About kit usage

What instruments, and equipment are required for the assay using this kit?
The instruments and equipment required for the use of this kit are listed below;
  • Deionized water (or distilled water)
  • Test tubes for preparation of standard solution series
  • Glassware for dilution of Wash Solution (10x) (a graduated cylinder, a bottle)
  • Pipettes (disposable tip type). One should be able to deliver 50 μL precisely, and another for 100 – 1,000 μL
  • Syringe-type repeating dispenser like Eppendorf multipette plus which can dispense 100 μL
  • Paper towel to remove washing buffer remaining in wells
  • A vortex-type mixer
  • A shaker for 96 well-plate (600 – 1,200 rpm)
  • An automatic washer for 96 well-plate (if available), or a wash bottle with a jet nozzle
  • A 96 well-plate reader (450 ± 10 nm, 620 nm :600 - 650 nm)
  • Software for data analysis
How should I use the human IL-7 standard?
The amount of purified water to be added varies by lot. Please reconstitute the human IL-7 standard with the amount of purified water specified in the “Reconstitution of standard” section on this product page to prepare the original standard solution (2,360 pg/mL). Then, dilute the solution using the kit-provided buffer that has been equilibrated to room temperature.
What are the storage conditions for each reagent?
[Buffer & TMB Solution]
Use only volume you need for your assay. Remaining reagents should be stored at 2-10℃ closing the cap tightly. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.
[Stop Solution]
Close the cap tightly and store at 2-10℃.
[Human IL-7 Standard & Biotin-conjugated Antibody]
The reconstituted standard solution (original standard solution) should be used within 2 weeks if stored in a refrigerator. Dispose the remaining diluted standard solutions after use.
[Peroxidase-conjugated Streptavidin Solution]
Remaining working solution (already diluted) should be disposed. The rest of the undiluted solution (unused): close the cap tightly and store at 2-10℃. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.

Prepare reagent solutions in appropriate volume for your assay.

Overview / Applications

Property

Manufacturer Information

Alias

  • 637-50441(previous code)
    AKH-IL7(previous code)

For research use or further manufacturing use only. Not for use in diagnostic procedures.

Product content may differ from the actual image due to minor specification changes etc.

If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.