Distributor
295-92701
Barcode No
4548995101846
60Tests
List Price
729.00 USD

Out of Stock

Distributor
291-92703
Barcode No
4548995101853
250Tests
List Price
1,948.00 USD

In stock in Japan

Document

SDS
Product Specification Sheet
Package Insert
Spectral Data
Certificate of Analysis
Calibration Certificate
Analytical Charts

Product Overview

Albumin is a plasma protein synthesized mainly in the liver, and measuring albumin concentration in serum and plasma is an important indicator in the diagnosis and management of liver disease and liver dysfunction. It is also known that in kidney diseases characterized by protein leakage into the urine, total protein and albumin concentrations in the blood decrease, causing hypoalbuminemia. Urinary albumin measurements are commonly performed using TIA method. Fujifilm Wako offers TIA reagents optimized for automated analyzers compatible with specimens from rats and mice.
This kit is a research reagent for an autoanalyzer to measure albumin in mouse urine or serum using anti-mouse albumin antibody. This is intended for research use only.

[Note] LBIS™ Mouse Urinary Albumin Assay Kits are research use only. They cannot be used for diagnostics purposes.

Kit Performance

Analysis sample Mouse Urine/Serum*
Calibration curve range 6.17-500 μg/mL
Sample volume 6 μL
Measurement duration Approx. 10 min.
* Since serum albumin concentrations are higher than those in urine, a dilution of approximately 100 times is required.

Kit Component

Components 60 tests 250 tests
1) Reagent 1 18.5 mL / 1 bottle 57.8 mL / 1 bottle
2) Reagent 2 6.6 mL / 1 bottle 16.5 mL / 1 bottle
3) Mouse Urinary Albumin Standard 2 mL / 1 bottle 1.5 mL / 2 bottles
4) Buffer for Standard 4 mL / 1 bottle 5 mL / 2 bottles

Data

Repeatability (within run precision)

Repeatability was conducted by measuring five-fold assays of serum samples from mouse at three concentrations with this product.

n\ID Mouse (BALB/c (6 weeks old ♂)) serum
A (μg/mL) B (μg/mL) C (μg/mL)
1 405 251 25.8
2 398 255 26.1
3 399 249 24.2
4 395 254 23.9
5 397 255 24.1
mean 399 253 24.8
SD 3.77 2.68 1.04
CV(%) 0.9 1.1 4.2

[Result]
The CV (%) of mouse serum was 0.9-4.2%, indicating good repeatability.

Reproducibility (between run precision)

Reproducibility was conducted by measuring serum samples from mouse at three concentrations each for four days with this product.

Day\ID Mouse (BALB/c (6 weeks old ♂)) serum
D (μg/mL) E (μg/mL) F (μg/mL)
1 255 128 20.5
2 249 120 20.1
3 250 122 21.9
4 251 125 20.3
mean 251 124 20.7
SD 2.63 3.50 0.82
CV(%) 1.0 2.8 3.9

[Result]
The CV (%) of mouse serum was 1.0-3.9%, indicating good reproducibility.

Dilution Linearity Test

Serum from mouse were serially diluted twofold with the buffer in this kit and measured in triplicate to assess linearity.

Mouse (BALB/c, 6 weeks old, ♂) serum

[Result]
All measurement results showed good linearity within the measurement range.

Spike and Recovery Test

Standard solutions at four concentrations were added to serum from mouse. Next, spike recovery tests were conducted, with each measurement performed in duplicate.

Mouse (BALB/c, 6 weeks old, ♂) sample

Spiked amount
(μg/mL)
Measurement value
(μg/mL)
Recovery volume
(μg/mL)
Recovery rate
(%)
Serum-1 0 15.7 - -
15 30.8 15.1 101
30 45.0 29.3 97.7
60 74.1 58.4 97.3
120 137 122 102
Average 99.5
Spiked amount
(μg/mL)
Measurement value
(μg/mL)
Recovery volume
(μg/mL)
Recovery rate
(%)
Serum-2 0 70.5 - -
100 174 104 104
150 221 151 101
200 268 198 99.0
250 316 246 98.4
Average 101

[Result]
Good recovery rates were confirmed.

FAQ

About sample

What should I do with samples that exceed the measurable range?
Dilute samples with purified water for an autoanalyzer, or dilution of samples should be made in a test tube using saline solution. Obtained value should be raised its dilution rate to be assay value.
Is it possible to conduct measurements using chylous samples or hemolyzed samples?
Hemolytic and chylous serum samples are not suitable.
* To avoid influence of blood (high lipid or hemolysis, etc.), if your original samples have heavy chyle or hemolysis, do not use them for assay. Abnormal value might be obtained with hemolysis above 200 mg/dL with this kit.

About kit usage

What instruments, and equipment are required for the assay using this kit?
  • An autoanalyzer
  • Purified water (distilled water)
  • Physiological saline
  • Test tubes for preparation of standard solution series
  • Pipettes (disposable tip type)
  • A vortex-type mixer.
What are the storage conditions guidelines for each reagent?
  1. Reagent 1
    Store at 2°C - 10°C.
  2. Reagent 2
    Store at 2°C - 10°C.
  3. Mouse Urinary Albumin Standard
    Store the unused reagents at 2°C - 10°C.
  4. Buffer for Standard
    Store at 2°C - 10°C.

Overview / Applications

Property

Manufacturer Information

Alias

  • AKRAL-021S(previous code)
    638-25561(previous code)
    AKRAL-021SZ1(previous code)
    634-25563(previous code)

For research use or further manufacturing use only. Not for use in diagnostic procedures.

Product content may differ from the actual image due to minor specification changes etc.

If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.