SPE Columns for PFAS Analysis
Fujifilm Wako offers the Presep® series of Solid-Phase Extraction (SPE) columns and cartridges for sample pretreatment in the analysis of foods, environmental, and pharmaceutical samples. Four products are particularly suitable for sample pretreatment in PFAS analysis: Presep® PFAS (60 mg/3 mL), Presep® PFAS (150 mg/6 mL), Presep®-C PFAS (Short), and Presep®-C Agri-II (Short).
Presep® PFAS is a mixed-mode column and cartridge packed with a reversed-phase polymer modified with weak anion-exchange groups, enabling efficient capture and extraction of PFAS across a wide range of carbon chain lengths. This enables highly efficient recovery of more than 20 types of PFAS.
Presep®-C Agri-II is a reversed-phase column and cartridge that efficiently recovers long-chain PFAS, including PFOS and PFOA. These products are useful for the extraction and concentration of PFAS from river water, drinking water, wastewater, and biological samples.
Presep® PFAS
Features
- Recovers a wide range of PFAS, from short-chain to long-chain
- Uses a divinylbenzene-polymethacrylate packing material modified with weak anion-exchange groups
- Recovers 24 PFAS compounds even at high flow rates (up to 10 mL/min)
Example of a Solid-Phase Extraction Procedure*
Cartridge type
Syringe type
Loading
Tap water (Add Internal standard solution (100 μL) and Standard solution (200 μL))
Flow rate: 5 mL/min
Washing
H2O (about 15 mL)
Dry (N2 purge)
Extraction (Backflush)
0.1% NH3 in CH3OH (5 mL)
Extraction (Forward Flow)
0.1% NH3 in CH3OH (5 mL)
Concentration
Nitrogen blowdown (→0.5 mL)
LC-MS(/MS) analysis
Injection volume: 5 μL
*This procedure is provided as an example. Conditions may need to be optimized depending on the sample.
Spike-and-Recovery Test of 10 PFAS in a Tap Water Sample Using Presep® PFAS*
Loading
Tap water (Add Internal standard solution (100 μL) and Standard solution (200 μL))
Loading volume: cartridge type 500 mL, syringe type 100 mL
Flow rate: 10 mL/min
Washing
CH3OH (15 mL)
Dry (N2 purge)
Extraction
0.1% NH3 in CH3OH (5 mL)
Concentration
Nitrogen blowdown (→1.0 mL)
If the volume is less than 1 mL, make up the volume to 1 mL with CH3OH.
LC-MS(/MS) analysis

| Compound Name・Recovery (%)* | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| C4 PFBA | ・106 | C5 PFPeA | ・101 | C6 PFHxA | ・102 | C7 PFHpA | ・111 | C8 PFOA | ・99 |
| C9 PFNA | ・106 | C4 PFBS | ・104 | C6 PFHxS | ・98 | C8 PFOS | ・99 | GenX | ・89 |

| Compound Name・Recovery (%)* | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| C4 PFBA | ・103 | C5 PFPeA | ・92 | C6 PFHxA | ・92 | C7 PFHpA | ・93 | C8 PFOA | ・94 |
| C9 PFNA | ・100 | C4 PFBS | ・91 | C6 PFHxS | ・89 | C8 PFOS | ・114 | GenX | ・101 |

| Compound Name・Recovery (%)* | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| C4 PFBA | ・124 | C5 PFPeA | ・118 | C6 PFHxA | ・123 | C7 PFHpA | ・126 | C8 PFOA | ・114 |
| C9 PFNA | ・109 | C4 PFBS | ・110 | C6 PFHxS | ・111 | C8 PFOS | ・109 | GenX | ・115 |
- These data are provided as examples only and do not guarantee product performance.
LC-MS(/MS) Conditions
[HPLC]
- Column
- Wakopak® Ultra C18-2 2.1 x 100 mm
- Delay Column
- Wakopak® Ultra C18-3 2.0 x 150 mm
- Column Temperature
- 40℃
- Eluent
- A) 0.01 mol/L CH3COONH4 in H2O
B) CH3CN - Gradient
-
Time (min) B conc. (%) 0-2 20 2-22 20-95 22-30 95 - Flow Rate
- 0.2 mL/min
[MS]
- Ionization
- ESI
- Mode
- MRM
Comparison of Spike-and-Recovery Tests for 30 PFAS Using Presep®-C PFAS (Short) and Company A’s Mixed-Mode Column*
PFAS recovery by the packing material was evaluated by adding the internal standard solution after extraction.
Loading
Sample (Add Standard solution (200 μL))
Loading volume: 500 mL
Flow rate: 10 mL/min
Washing
CH3OH (15 mL)
Dry (N2 purge)
Extraction (Backflush)
0.1% NH3 in CH3OH (5 mL)
Concentration
Nitrogen blowdown (→0.7 mL)
Add Internal standard solution (100 μL)
If the volume is less than 1 mL, make up the volume to 1 mL CH3OH.
LC-MS(/MS) analysis

| Compound Name・Recovery (%)* | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| C4 PFBA | ・90/90 | C5 PFPeA | ・102/92 | C6 PFHxA | ・102/87 | C7 PFHpA | ・106/89 | C8 PFOA | ・96/90 |
| C9 PFNA | ・106/90 | C10 PFDA | ・106/88 | C11 PFUnDA | ・101/86 | C12 PFDoDA | ・94/79 | C13 PFTrDA | ・84/71 |
| C14 PFTeDA | ・85/70 | C16 PFHxDA | ・88/84 | C18 PFODA | ・103/84 | C4 PFBS | ・103/83 | C5 PFPeS | ・94/85 |
| C6 PFHxS | ・92/86 | C7 PFHpS | ・111/101 | C8 PFOS | ・92/84 | C10 PFDS | ・92/83 | C6 4:2 FTS | ・111/87 |
| C8 6:2 FTS | ・101/90 | C10 8:2 FTS | ・101/89 | C8 PFOSA | ・8/10 | C8 N-MeFOSA | ・0/1 | C8 N-EtFOSAA | ・88/78 |
| GenX | ・104/99 | PFMPA | ・117/105 | PFMBA | ・86/70 | PFEESA | ・103/108 | C10 8:2 FTUCA | ・99/86 |

| Compound Name・Recovery (%)* | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| C4 PFBA | ・80/109 | C5 PFPeA | ・106/106 | C6 PFHxA | ・94/90 | C7 PFHpA | ・88/96 | C8 PFOA | ・96/98 |
| C9 PFNA | ・98/96 | C10 PFDA | ・84/83 | C11 PFUnDA | ・84/84 | C12 PFDoDA | ・78/77 | C13 PFTrDA | ・66/65 |
| C14 PFTeDA | ・71/74 | C16 PFHxDA | ・72/77 | C18 PFODA | ・57/57 | C4 PFBS | ・94/92 | C5 PFPeS | ・94/91 |
| C6 PFHxS | ・103/94 | C7 PFHpS | ・100/102 | C8 PFOS | ・97/99 | C10 PFDS | ・86/89 | C6 4:2 FTS | ・107/101 |
| C8 6:2 FTS | ・99/97 | C10 8:2 FTS | ・83/84 | C8 PFOSA | ・81/13 | C8 N-MeFOSA | ・0/0 | C8 N-EtFOSAA | ・75/71 |
| GenX | ・97/88 | PFMPA | ・78/74 | PFMBA | ・99/94 | PFEESA | ・92/78 | C10 8:2 FTUCA | ・80/53 |
- These data are provided as examples only and do not guarantee product performance.
LC-MS(/MS) Conditions
[HPLC]
- Column
- Wakopak® Ultra C18-2 2.1 x 100 mm
- Delay Column
- Wakopak® Ultra C18-3 2.0 x 150 mm
- Column Temperature
- 40℃
- Eluent
- A) 0.01 mol/L CH3COONH4 in H2O
B) CH3CN - Gradient
-
Time (min) B conc. (%) 0-2 20 2-22 20-95 22-30 95 30-40 20 - Flow Rate
- 0.2 mL/min
[MS]
- Ionization
- ESI
- Mode
- MRM
Spike-and-Recovery Test of 28 PFAS in a River Water Sample Using Presep®-C PFAS (Short)*
Loading
River water (Add Internal standard solution (100 μL) and Standard solution (200 μL))
Loading volume: cartridge type 500 mL
Flow rate: 10 mL/min
Washing
CH3OH (15 mL)
Dry (N2 purge)
Extraction (Backflush)
0.1% NH3 in CH3OH (5 mL)
Concentration
Nitrogen blowdown (→0.7 mL)
If the volume is less than 1 mL, make up the volume to 1 mL with CH3OH.
LC-MS(/MS) analysis

| Compound Name・Recovery (%)* | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| C4 PFBA | ・108 | C5 PFPeA | ・99 | C6 PFHxA | ・97 | C7 PFHpA | ・97 | C8 PFOA | ・102 |
| C9 PFNA | ・101 | C10 PFDA | ・100 | C11 PFUnDA | ・107 | C12 PFDoDA | ・110 | C13 PFTrDA | ・111 |
| C14 PFTeDA | ・112 | C16 PFHxDA | ・94 | C18 PFODA | ・105 | C4 PFBS | ・99 | C5 PFPeS | ・101 |
| C6 PFHxS | ・101 | C7 PFHpS | ・113 | C8 PFOS | ・123 | C10 PFDS | ・95 | C6 4:2 FTS | ・98 |
| C8 6:2 FTS | ・96 | C10 8:2 FTS | ・101 | C8 N-EtFOSAA | ・100 | GenX | ・101 | PFMPA | ・122 |
| PFMBA | ・117 | PFEESA | ・107 | C10 8:2 FTUCA | ・84 | ||||
- These data are provided as examples only and do not guarantee product performance.
Presep®-C Agri-II (Short)
Features
- Cartridge-type reverse phase solid-phase extraction column using a styrene-divinylbenzene-methacrylate polymer as the packing material
- Capable of efficiently recovering long-chain PFAS, including PFOS and PFOA
Spike-and-Recovery Test for PFOA and PFOS in a Tap Water Sample Using Presep®-C Agri-II (Short)*
Loading
Tap water (Add Internal standard solution (50 μL) and Standard solution (100 μL))
Loading volume: 500 mL
Flow rate: 5 mL/min
Washing
H2O (15 mL)
Dry (N2 purge)
Extraction (Backflush)
0.1% NH3 in CH3OH (5 mL)
Concentration
Nitrogen blowdown (→0.5 mL)
LC-MS(/MS) analysis
LC-MS(/MS) Conditions
- Column
- Wakopak® Ultra C18-3 2.0 mm x 150 mm
- Eluent
- A) 0.01 mol/L CH3COONH4 in H2O
B) CH3CN - LC-MS/MS
- Prominence LC-20ADXR / LCMS-8030 plus (Shimadzu Corporation)
- Gradient
-
Time (min) B conc. (%) 0-1 25 1-26 25-100 26-30 100 30-30.01 100-25 30.01-35 25
Measurement Results
Chromatogram

Recovery Results
| Detected ion | Recovery | ||
|---|---|---|---|
| Precursor | Product | ||
| PFOS | 498.8 | 80.0 | 100.9% |
| PFOA | 412.9 | 169.1 | 95.0% |
- These data are provided as examples only and do not guarantee product performance.
Example of Monitoring Ions
| Compound Name | Carbon Chain | Detected ion (m/z) | Mode | |
|---|---|---|---|---|
| Precursor | Product | |||
| PFBA | 4 | 212.95 | 169.00 | - |
| PFPeA | 5 | 262.95 | 219.10 | - |
| PFHxA | 6 | 313.00 | 268.95 | - |
| PFHpA | 7 | 363.00 | 319.00 | - |
| PFOA | 8 | 413.00 | 168.95 | - |
| PFNA | 9 | 463.00 | 419.00 | - |
| PFDA | 10 | 512.95 | 469.00 | - |
| PFUnDA | 11 | 563.00 | 518.90 | - |
| PFDoDA | 12 | 612.95 | 569.00 | - |
| PFTrDA | 13 | 662.95 | 618.95 | - |
| PFTeDA | 14 | 712.95 | 668.80 | - |
| PFHxDA | 16 | 812.95 | 768.85 | - |
| PFODA | 18 | 912.70 | 868.95 | - |
| PFBS | 4 | 298.95 | 80.00 | - |
| PFPeS | 5 | 349.00 | 80.05 | - |
| PFHxS | 6 | 398.95 | 80.00 | - |
| PFHpS | 7 | 449.00 | 80.10 | - |
| PFOS | 8 | 499.00 | 80.15 | - |
| PFDS | 10 | 598.95 | 99.10 | - |
| 4:2 FTS | 6 | 326.95 | 306.90 | - |
| 6:2 FTS | 8 | 427.00 | 407.05 | - |
| 8:2 FTS | 10 | 527.00 | 506.85 | - |
| PFOSA | 8 | 498.00 | 77.95 | - |
| N-MeFOSA | 8 | 511.95 | 169.05 | - |
| N--EtFOSAA | 8 | 584.00 | 418.90 | - |
| GenX | - | 284.90 | 169.05 | - |
| PFMPA | - | 229.00 | 85.00 | - |
| PFMBA | - | 278.95 | 85.05 | - |
| PFEESA | - | 314.95 | 135.00 | - |
| 8:2 FTUCA | 10 | 457.00 | 392.95 | - |
Product List