Columns for analysis of DNPH-derivatized aldehydes and ketones

HPLC Column for Environment Analysis

Wakopak® Wakosil® DNPH, Wakopak® Wakosil® DNPH-Ⅱ

These are columns for the HPLC analysis of aldehydes and ketones in the atmosphere. These special columns are optimized for the simultaneous analysis of 2,4-dinitrophenylhydrazone (DNPH)-derivatized aldehydes and ketones. Aldehydes are subject to regulation and measurement as hazardous air pollutants under the Air Pollution Control Act, Offensive Odor Control Law, etc.​

​Wakopak® Wakosil® DNPH can separate n- and iso-butyraldehydes, which are difficult to separate with general HPLC columns. It can analyze the same components as those analyzed by GC.

Wakopak® Wakosil® DNPH-II can separate 15 components regulated by the US EPA within 20 minutes.​

Example of analysis of 16 DNPH aldehydes

Wakopak® Wakosil® DNPH

n- and iso-butyraldehydes can be separated by HPLC.

Column :Wakopak® Wakosil® DNPH (4.6 mm x 250 mm) Sample: 1) Formaldehyde-2,4-DNPH
2) Acetaldehyde-2,4-DNPH
3) Propionaldehyde-2,4-DNPH
4) Acrolein-2,4-DNPH
5) Acetone-2,4-DNPH
6) iso-Butyraldehyde-2,4-DNPH
7) n-Butyraldehyde-2,4-DNPH
8) Crotonaldehyde-2,4-DNPH
9) iso-Valeraldehyde-2,4-DNPH
10) n-Valeraldehyde-2,4-DNPH
11) Benzaldehyde-2,4-DNPH
12) Hexaldehyde-2,4-DNPH
13) o-Tolualdehyde-2,4-DNPH
14) m-Tolualdehyde-2,4-DNPH
15) p-Tolualdehyde-2,4-DNPH
16) 2,5-Dimethylbenzaldehyde-2,4-DNPH
each 0.625 µg/mL (as aldehyde, ketone)
Eluent :A) Wakosil® DNPH Eluent A
 B) Wakosil® DNPH Eluent B
 0 - 16 min. B 10 %
 16 - 35 min. B 10 - 90 %
 35 - 45 min. B 90 %
Flow Rate :0.6 mL/min. at 37 ℃
Detection :UV 360 nm, 0.016 AUFS

Analysis of 15 DNPH aldehydes (rapid analysis)

Wakopak® Wakosil® DNPH-II

1. Blank

One Presep®-C DNPH cartridge eluted with acetonitrile and diluted to 5 mL.

2. Example of analysis of indoor air sample

30 L of air in the laboratory sampled at a flow rate of 0.5 L/min using Presep®-C DNPH and eluted in the same manner as described above.​

3. Reference standard ​

Chromatogram of the reference standard of 15 DNPH-derivatized aldehydes​

  • Column :Wakopak® Wakosil® DNPH-II (4.6 mm x 150 mm)
    Eluent :A) Wakosil® DNPH-Ⅱ Eluent A
     B)Wakosil® DNPH-Ⅱ Eluent B
     0 - 10 min  B conc. 0-100 %
     10 - 16 min B conc. 100 %
    Flow Rate :1.0 mL/min. at 35 ℃
    Detection :UV 360 nm
    Injection Volume :10 µL
  • Sample: 1) Formaldehyde-2,4-DNPH
    2) Acetaldehyde-2,4-DNPH
    3) Acetone-2,4-DNPH
    4) Acrolein-2,4-DNPH
    5) Propionaldehyde-2,4-DNPH
    6) Crotonaldehyde-2,4-DNPH
    7) n-,iso-Butyraldehyde-2,4-DNPH
    8) Benzaldehyde-2,4-DNPH
    9) iso-Valeraldehyde-2,4-DNPH
    10) n-Valeraldehyde-2,4-DNPH
    11) o-Tolualdehyde-2,4-DNPH
    12) m-Tolualdehyde-2,4-DNPH
    13) p-Tolualdehyde-2,4-DNPH
    14) Hexaldehyde-2,4-DNPH
    15) 2,5-Dimethylbenzaldehyde-2,4-DNPH
    each 0.1 μg/mL (as aldehyde,ketone)

Product List

  • Open All
  • Close All

Column for HPLC

DNPH Eluent

Mixture Standard Solution

Solid-phase extraction column for pretreatment

Solvent

For research use or further manufacturing use only. Not for use in diagnostic procedures.

Product content may differ from the actual image due to minor specification changes etc.