SPE Columns for PFAS Analysis

Fujifilm Wako offers the Presep® series of Solid-Phase Extraction (SPE) columns and cartridges for sample pretreatment in the analysis of foods, environmental, and pharmaceutical samples. Four products are particularly suitable for sample pretreatment in PFAS analysis: Presep® PFAS (60 mg/3 mL), Presep® PFAS (150 mg/6 mL), Presep®-C PFAS (Short), and Presep®-C Agri-II (Short).
Presep® PFAS is a mixed-mode column and cartridge packed with a reversed-phase polymer modified with weak anion-exchange groups, enabling efficient capture and extraction of PFAS across a wide range of carbon chain lengths. This enables highly efficient recovery of more than 20 types of PFAS.
Presep®-C Agri-II is a reversed-phase column and cartridge that efficiently recovers long-chain PFAS, including PFOS and PFOA. These products are useful for the extraction and concentration of PFAS from river water, drinking water, wastewater, and biological samples.

Presep® PFAS

Features

  • Recovers a wide range of PFAS, from short-chain to long-chain
  • Uses a divinylbenzene-polymethacrylate packing material modified with weak anion-exchange groups
  • Recovers 24 PFAS compounds even at high flow rates (up to 10 mL/min)

Example of a Solid-Phase Extraction Procedure*

Cartridge type

Syringe type

Column Conditioning
1) 0.1% NH3 in CH3OH (4 mL)
2) CH3OH (4 mL)
3) H2O (4 mL)

Loading
Tap water (Add Internal standard solution (100 μL) and Standard solution (200 μL))
Flow rate: 5 mL/min
Washing
H2O (about 15 mL)

Dry (N2 purge)

Extraction (Backflush)
0.1% NH3 in CH3OH (5 mL)

Extraction (Forward Flow)
0.1% NH3 in CH3OH (5 mL)

Concentration
Nitrogen blowdown (→0.5 mL)

LC-MS(/MS) analysis
Injection volume: 5 μL

*This procedure is provided as an example. Conditions may need to be optimized depending on the sample.

Spike-and-Recovery Test of 10 PFAS in a Tap Water Sample Using Presep® PFAS*

Column Conditioning
1) 0.1% NH3 in CH3OH (4 mL)
2) CH3OH (4 mL)
3) H2O (4 mL)

Loading
Tap water (Add Internal standard solution (100 μL) and Standard solution (200 μL))
Loading volume: cartridge type 500 mL, syringe type 100 mL
Flow rate: 10 mL/min
Washing
CH3OH (15 mL)

Dry (N2 purge)

Extraction
0.1% NH3 in CH3OH (5 mL)

Concentration
Nitrogen blowdown (→1.0 mL)
If the volume is less than 1 mL, make up the volume to 1 mL with CH3OH.

LC-MS(/MS) analysis

Compound Name・Recovery (%)*
C4 PFBA ・106 C5 PFPeA ・101 C6 PFHxA ・102 C7 PFHpA ・111 C8 PFOA ・99
C9 PFNA ・106 C4 PFBS ・104 C6 PFHxS ・98 C8 PFOS ・99 GenX ・89
Compound Name・Recovery (%)*
C4 PFBA ・103 C5 PFPeA ・92 C6 PFHxA ・92 C7 PFHpA ・93 C8 PFOA ・94
C9 PFNA ・100 C4 PFBS ・91 C6 PFHxS ・89 C8 PFOS ・114 GenX ・101
Compound Name・Recovery (%)*
C4 PFBA ・124 C5 PFPeA ・118 C6 PFHxA ・123 C7 PFHpA ・126 C8 PFOA ・114
C9 PFNA ・109 C4 PFBS ・110 C6 PFHxS ・111 C8 PFOS ・109 GenX ・115
  • These data are provided as examples only and do not guarantee product performance.​

LC-MS(/MS) Conditions

[HPLC]

Column
Wakopak® Ultra C18-2 2.1 x 100 mm
Delay Column
Wakopak® Ultra C18-3 2.0 x 150 mm
Column Temperature
40℃
Eluent
A) 0.01 mol/L CH3COONH4 in H2O
B) CH3CN
Gradient
Time (min) B conc. (%)
0-2 20
2-22 20-95
22-30 95
Flow Rate
0.2 mL/min

[MS]

Ionization
ESI
Mode
MRM

Comparison of Spike-and-Recovery Tests for 30 PFAS Using Presep®-C PFAS (Short) and Company A’s Mixed-Mode Column*

PFAS recovery by the packing material was evaluated by adding the internal standard solution after extraction.​

Column Conditioning
1) 0.1% NH3 in CH3OH (4 mL)
2) CH3OH (4 mL)
3) H2O (4 mL)

Loading
Sample (Add Standard solution (200 μL))
Loading volume: 500 mL
Flow rate: 10 mL/min
Washing
CH3OH (15 mL)

Dry (N2 purge)

Extraction (Backflush)
0.1% NH3 in CH3OH (5 mL)

Concentration
Nitrogen blowdown (→0.7 mL)
Add Internal standard solution (100 μL)
If the volume is less than 1 mL, make up the volume to 1 mL CH3OH.

LC-MS(/MS) analysis

Compound Name・Recovery (%)*
C4 PFBA ・90/90 C5 PFPeA ・102/92 C6 PFHxA ・102/87 C7 PFHpA ・106/89 C8 PFOA ・96/90
C9 PFNA ・106/90 C10 PFDA ・106/88 C11 PFUnDA ・101/86 C12 PFDoDA ・94/79 C13 PFTrDA ・84/71
C14 PFTeDA ・85/70 C16 PFHxDA ・88/84 C18 PFODA ・103/84 C4 PFBS ・103/83 C5 PFPeS ・94/85
C6 PFHxS ・92/86 C7 PFHpS ・111/101 C8 PFOS ・92/84 C10 PFDS ・92/83 C6 4:2 FTS ・111/87
C8 6:2 FTS ・101/90 C10 8:2 FTS ・101/89 C8 PFOSA ・8/10 C8 N-MeFOSA ・0/1 C8 N-EtFOSAA ・88/78
GenX ・104/99 PFMPA ・117/105 PFMBA ・86/70 PFEESA ・103/108 C10 8:2 FTUCA ・99/86
Compound Name・Recovery (%)*
C4 PFBA ・80/109 C5 PFPeA ・106/106 C6 PFHxA ・94/90 C7 PFHpA ・88/96 C8 PFOA ・96/98
C9 PFNA ・98/96 C10 PFDA ・84/83 C11 PFUnDA ・84/84 C12 PFDoDA ・78/77 C13 PFTrDA ・66/65
C14 PFTeDA ・71/74 C16 PFHxDA ・72/77 C18 PFODA ・57/57 C4 PFBS ・94/92 C5 PFPeS ・94/91
C6 PFHxS ・103/94 C7 PFHpS ・100/102 C8 PFOS ・97/99 C10 PFDS ・86/89 C6 4:2 FTS ・107/101
C8 6:2 FTS ・99/97 C10 8:2 FTS ・83/84 C8 PFOSA ・81/13 C8 N-MeFOSA ・0/0 C8 N-EtFOSAA ・75/71
GenX ・97/88 PFMPA ・78/74 PFMBA ・99/94 PFEESA ・92/78 C10 8:2 FTUCA ・80/53
  • These data are provided as examples only and do not guarantee product performance.

LC-MS(/MS) Conditions

[HPLC]

Column
Wakopak® Ultra C18-2 2.1 x 100 mm
Delay Column
Wakopak® Ultra C18-3 2.0 x 150 mm
Column Temperature
40℃
Eluent
A) 0.01 mol/L CH3COONH4 in H2O
B) CH3CN
Gradient
Time (min) B conc. (%)
0-2 20
2-22 20-95
22-30 95
30-40 20
Flow Rate
0.2 mL/min

[MS]

Ionization
ESI
Mode
MRM

Spike-and-Recovery Test of 28 PFAS in a River Water Sample Using Presep®-C PFAS (Short)*

Column Conditioning
1) 0.1% NH3 in CH3OH (4 mL)
2) CH3OH (4 mL)
3) H2O (4 mL)

Loading
River water (Add Internal standard solution (100 μL) and Standard solution (200 μL))
Loading volume: cartridge type 500 mL
Flow rate: 10 mL/min
Washing
CH3OH (15 mL)

Dry (N2 purge)

Extraction (Backflush)
0.1% NH3 in CH3OH (5 mL)

Concentration
Nitrogen blowdown (→0.7 mL)
If the volume is less than 1 mL, make up the volume to 1 mL with CH3OH.

LC-MS(/MS) analysis

Compound Name・Recovery (%)*
C4 PFBA ・108 C5 PFPeA ・99 C6 PFHxA ・97 C7 PFHpA ・97 C8 PFOA ・102
C9 PFNA ・101 C10 PFDA ・100 C11 PFUnDA ・107 C12 PFDoDA ・110 C13 PFTrDA ・111
C14 PFTeDA ・112 C16 PFHxDA ・94 C18 PFODA ・105 C4 PFBS ・99 C5 PFPeS ・101
C6 PFHxS ・101 C7 PFHpS ・113 C8 PFOS ・123 C10 PFDS ・95 C6 4:2 FTS ・98
C8 6:2 FTS ・96 C10 8:2 FTS ・101 C8 N-EtFOSAA ・100 GenX ・101 PFMPA ・122
PFMBA ・117 PFEESA ・107 C10 8:2 FTUCA ・84
  • These data are provided as examples only and do not guarantee product performance.

Presep®-C Agri-II (Short)

Features

  • Cartridge-type reverse phase solid-phase extraction column using a styrene-divinylbenzene-methacrylate polymer as the packing material
  • Capable of efficiently recovering long-chain PFAS, including PFOS and PFOA

Spike-and-Recovery Test for PFOA and PFOS in a Tap Water Sample Using Presep®-C Agri-II (Short)*

Column Conditioning
1) 0.1% NH3 in CH3OH (4 mL)
2) CH3OH (4 mL)
3) H2O (4 mL)

Loading
Tap water (Add Internal standard solution (50 μL) and Standard solution (100 μL))
Loading volume: 500 mL
Flow rate: 5 mL/min
Washing
H2O (15 mL)

Dry (N2 purge)

Extraction (Backflush)
0.1% NH3 in CH3OH (5 mL)

Concentration
Nitrogen blowdown (→0.5 mL)

LC-MS(/MS) analysis

LC-MS(/MS) Conditions

Column
Wakopak® Ultra C18-3 2.0 mm x 150 mm
Eluent
A) 0.01 mol/L CH3COONH4 in H2O
B) CH3CN
LC-MS/MS
Prominence LC-20ADXR / LCMS-8030 plus (Shimadzu Corporation)
Gradient
Time (min) B conc. (%)
0-1 25
1-26 25-100
26-30 100
30-30.01 100-25
30.01-35 25

Measurement Results

Chromatogram
Recovery Results
Detected ion Recovery
Precursor Product
PFOS 498.8 80.0 100.9%
PFOA 412.9 169.1 95.0%
  • These data are provided as examples only and do not guarantee product performance.

Example of Monitoring Ions

Compound Name Carbon Chain Detected ion (m/z) Mode
Precursor Product
PFBA 4 212.95 169.00 -
PFPeA 5 262.95 219.10 -
PFHxA 6 313.00 268.95 -
PFHpA 7 363.00 319.00 -
PFOA 8 413.00 168.95 -
PFNA 9 463.00 419.00 -
PFDA 10 512.95 469.00 -
PFUnDA 11 563.00 518.90 -
PFDoDA 12 612.95 569.00 -
PFTrDA 13 662.95 618.95 -
PFTeDA 14 712.95 668.80 -
PFHxDA 16 812.95 768.85 -
PFODA 18 912.70 868.95 -
PFBS 4 298.95 80.00 -
PFPeS 5 349.00 80.05 -
PFHxS 6 398.95 80.00 -
PFHpS 7 449.00 80.10 -
PFOS 8 499.00 80.15 -
PFDS 10 598.95 99.10 -
4:2 FTS 6 326.95 306.90 -
6:2 FTS 8 427.00 407.05 -
8:2 FTS 10 527.00 506.85 -
PFOSA 8 498.00 77.95 -
N-MeFOSA 8 511.95 169.05 -
N--EtFOSAA 8 584.00 418.90 -
GenX - 284.90 169.05 -
PFMPA - 229.00 85.00 -
PFMBA - 278.95 85.05 -
PFEESA - 314.95 135.00 -
8:2 FTUCA 10 457.00 392.95 -

Product List

  • Open All
  • Close All

Solid-Phase Extraction (SPE) Columns

Related Product List

  • Open All
  • Close All

HPLC Columns

Solvents for PFAS Analysis

Mixture Standard Solutions

Mixture Internal Standard Solutions

Analytical Standards

For research use or further manufacturing use only. Not for use in diagnostic procedures.

Product content may differ from the actual image due to minor specification changes etc.

If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.

+49-2131-311-271