Wakopak® Core C18 ADRA
This product is a column suitable for the analysis of ADRA(OECD TG442C). It contains core-shell filler of 2.6μm in particle diameter.
It is checked for each column that NAC, NAL and other chemicals are respectively separated.
Column Performance Report
Analysis Examples
Analysis examples of 10 proficiency substances recommended in OECD Test guideline.
Green:Chemical
- p-Benzoquinone
- Sodium p-Toluenesulfonchloramide Trihydrate
- Cinnamaldehyde
- Palmitoyl Chloride
- Imidazolidinyl Urea
- Famesal
- Glycerin
- Benzyl Alcohol
- Dimethyl Isophthalate
- Propyl p-Hydroxybenzoate
Green:Chemical
HPLC Conditions
Instrument | Agilent UHPLC 1290 infinity Ⅱ (Ailent) |
---|---|
Column | Wakopak® Core C18 ADRA 3.0 x 150 mm (D) |
Mobile phase A | 0.1 % Trifluoroacetic Acid-Water |
Mobile phase B | 0.1 % Trifluoroacetic Acid-Acetonitrile |
NAC Gradient conditions | 0min.(A:B=70:30)→9.5min.(45:55)→10.0min(0:100) →13.0min.(0:100)→13.5min.(70:30)→25.0min.(70:30) |
NAL Gradient conditions | 0min.(A:B=80:20)→9.5min.(55:45)→10.0min(0:100)→13.0min.(0:100)→13.5min.(80:20)→25.0min.(80:20) |
Flow rate | 0.3 mL/min |
Mixer volume | 35 μL |
Column temperature | 40 ℃ |
UV detector | UV 281 nm |
Injection Quantity | 10 μL |