Edman Method (Protein Sequencer)
The Edman method is an excellent and widely adopted method for determining the amino acid sequence in proteins and peptides. The method identifies amino acid residues one by one from the N-terminal side.
Since the automated Edman method was developed by Edman and Begg, microanalysis and acceleration have been required, and analyses in the order of pmol have been performed using a sequencer(PPSQ).
Edman Degradation Method
Edman degradation method is a method to determine the primary structure of proteins by sequentially cleaving peptide bonds from the N-terminus of proteins or peptides.
Principle
- Phenyl isothiocyanate (PITC) should be reacted with free amino groups at the N-terminus of proteins or peptides to form an phenylthiocarbamyl amino acid (PTC-amino acid).
- Isolate an anilinothiazolinone amino acid (ATZ-amino acid) using trifluoroacetic acid (TFA).
- Convert to a stable 3-phenyl-2-thiohydantoic acid (PTH-amino acid) under acidic conditions for analysis.
Analytical Procedures
In 1967, Edman developed an automated processing device for Edman degradation in a liquid phase2). In the 1980s, the first Japanese gas-phase protein sequencer PSQ-1 was developed to automatically perform Edman degradation in the gas phase and analyze amino acids via isocratic elution with high-speed liquid chromatography (HPLC)3). Subsequently, the PPSQ (protein sequencer) series have evolved through the development of analyzing unit and data processing software.
The flow chart of automated amino acid sequencing analysis is shown in Fig 1.
Automated amino acid sequencing analysis begins with the development of Edman degradation in the reactor and converter. The stable PTH-amino acid is injected into HPLC for identification, and the data are analyzed in the data processing unit.
For PPSQ, the isocratic elution with a constant mobile-phase composition and the gradient elution in which the mobile phase composition is continuously changing are available.
The chromatogram for the gradient elution of a PTH-amino acid mixture standard (2 pmol) is shown in Fig 2 (a), and the chromatogram for the isocratic elution is shown in Fig 2 (b).
Overall, gradient elution enables the detection of peaks that are about 3 to 5 times higher than those of isocratic elution, allowing the separation of highly hydrophobic amino acids.
In addition, the gradient elution enables PTH-amino acid separation and detection even when trace 500 fmol samples are analyzed and enables easy detection of increasing PTH-amino acids (Fig 3).
For PPSQ, various reaction reagents and eluents are required, and the processing is complicated. “Reagents for SHIMADZU Protein Peptide Sequencer” sold in FUJIFILM Wako are dedicated for PPSQ, of which container can be directly set on the equipment.
Background Reduction
The background that appears on the chromatogram includes dimethyl phenylthiourea (DMPTU), diphenylthiourea (DPTU), and diphenylurea (DPU), which are mainly by-products of Edman degradation, and substances contained in Edman’s reagent. These are produced from phenyl isothiocyanate (PITC); a coupling reagent; and trimethylamine (TMA), which creates a basic atmosphere. When the quantity of sample is small, the peaks of DMTPU; this by-product; and dithiothreitol (DTT), an antioxidant contained in 25% trifluoroacetic acid used for conversion reaction, interfere with the detection of amino acid peaks (PTH-Asp, PTH-His).
These problems can be solved by changing the reagent that creates a basic atmosphere from TMA to N-methylpiperidine and using 25% trifluoroacetic acid, which does not contain DTT (Fig 4).
References
- Kuriki, T. et al. : Wako Jun-yaku Jiho, 86 (4), 6 (2018).
- Edman, P. and Begg, G. : Eur. J. Bouchem., 1,80 (1967).
- Ueda, A. et al.: SHIMADU Hyouron, 46 (4), 207 (1989).
Product List
- Open All
- Close All
PTH-Amino Acid Standard
SHIMADZU Protein Peptide Sequencer
Dedicated Eluent
For research use or further manufacturing use only. Not for use in diagnostic procedures.
Product content may differ from the actual image due to minor specification changes etc.
If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.