Wakopak® Core C18 ADRA
This product is a column suitable for the analysis of ADRA(OECD TG442C). It contains core-shell filler of 2.6μm in particle diameter.
It is checked for each column that NAC, NAL and other chemicals are respectively separated.
Column Performance Report
Analysis Examples
Analysis examples of 10 proficiency substances recommended in OECD Test guideline.
Green:Chemical
- p-Benzoquinone
- Sodium p-Toluenesulfonchloramide Trihydrate
- Cinnamaldehyde
- Palmitoyl Chloride
- Imidazolidinyl Urea
- Famesal
- Glycerin
- Benzyl Alcohol
- Dimethyl Isophthalate
- Propyl p-Hydroxybenzoate
Green:Chemical
HPLC Conditions
Instrument | Agilent UHPLC 1290 infinity Ⅱ (Ailent) |
---|---|
Column | Wakopak® Core C18 ADRA 3.0 x 150 mm (D) |
Mobile phase A | 0.1 % Trifluoroacetic Acid-Water |
Mobile phase B | 0.1 % Trifluoroacetic Acid-Acetonitrile |
NAC Gradient conditions | 0min.(A:B=70:30)→9.5min.(45:55)→10.0min(0:100) →13.0min.(0:100)→13.5min.(70:30)→25.0min.(70:30) |
NAL Gradient conditions | 0min.(A:B=80:20)→9.5min.(55:45)→10.0min(0:100)→13.0min.(0:100)→13.5min.(80:20)→25.0min.(80:20) |
Flow rate | 0.3 mL/min |
Mixer volume | 35 μL |
Column temperature | 40 ℃ |
UV detector | UV 281 nm |
Injection Quantity | 10 μL |
The details of ADRA Kit
Product List
- Open All
- Close All
For research use or further manufacturing use only. Not for use in diagnostic procedures.
Product content may differ from the actual image due to minor specification changes etc.
If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.
The prices are list prices in Japan.Please contact your local distributor for your retail price in your region.