Groundwater Control Analytical Standard

Standard samples of nonylphenol, 1,4-dioxane, and1,2-dichloroethylene used for Environmental Quality Standards for the Water Pollution in Base Environment Act in Japan are available. Nonylphenols are suspected to be endocrine disrupting chemicals. The quantitative analysis for it is carried out by solid phase extraction - GC/MS SIM method. 4-(3,6-dimethlyheputane-3yl)phonol-13C6, a related product, is also available.

Application

GC-MS chart (total ion chromatogram, TIC)

00348_img01.png

Conditions

GC Column BPX-5 made by SGE (0.25φmm x 30m, thickness: 0.25μm)
Temperature 50℃ (1min) → 8℃/min → 300℃
Vaporizing temperature 250℃
Carrier He, 1.2 mL/min
Injection mode Spiritless
MS Ionization mode EI
Interface temperature 280℃
Ion Source 230℃
Sample amount 2.5 mg/mL (CH3COCH3), 1 μL


Enlarged chromatograph (mixture of nonylphenol isomers)

00348_img02.jpg

Product List

  • Open All
  • Close All