Groundwater Control Analytical Standard

Standard samples of nonylphenol, 1,4-dioxane, and1,2-dichloroethylene used for Environmental Quality Standards for the Water Pollution in Base Environment Act in Japan are available. Nonylphenols are suspected to be endocrine disrupting chemicals. The quantitative analysis for it is carried out by solid phase extraction - GC/MS SIM method. 4-(3,6-dimethlyheputane-3yl)phonol-13C6, a related product, is also available.

Application

GC-MS chart (total ion chromatogram, TIC)

00348_img01.png

Conditions

GC Column BPX-5 made by SGE (0.25φmm x 30m, thickness: 0.25μm)
Temperature 50℃ (1min) → 8℃/min → 300℃
Vaporizing temperature 250℃
Carrier He, 1.2 mL/min
Injection mode Spiritless
MS Ionization mode EI
Interface temperature 280℃
Ion Source 230℃
Sample amount 2.5 mg/mL (CH3COCH3), 1 μL


Enlarged chromatograph (mixture of nonylphenol isomers)

00348_img02.jpg

Product List

  • Open All
  • Close All

Nonylphenol

Vinyl chloride (chloroethene)

1,4-dioxane

1,2-Dichloroethene

Related Product List

  • Open All
  • Close All

Nonylphenol internal standard / Surrogate

Analysis column

Author: E. Leung

For research use or further manufacturing use only. Not for use in diagnostic procedures.

Product content may differ from the actual image due to minor specification changes etc.

If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.

The prices are list prices in Japan.Please contact your local distributor for your retail price in your region.

Please contact us via the inquiry form.