It has been added to the cart.

It has been added to the comparison table.

LabAssay™ LDL-Cholesterol

for Cellbiology
Manufacturer :
FUJIFILM Wako Pure Chemical Corporation
Storage Condition :
Keep at 2-10 degrees C.
  • Structural Formula
  • Label
  • Packing
SDS
Comparison
Product Number
Package Size
Price
Inventory
Distributor
291-96701
Barcode No
4548995099990
100Tests
Inquire

In stock in Japan

Document

SDS ...all
Product Specification Sheet
Package Insert
Spectral Data
Certificate of Analysis
Calibration Certificate

Kit component

100 tests

Pretreatment 20 mL
Reacting Solution 8 mL
LDL-Cholesterol Standard 2 vials
Standard Diluent 10 mL

Product Overview

04572869_img01.jpg

The cholesterol carried by Low Density Lipoprotein (LDL) termed LDL-cholesterol. LDL carries cholesterol produced in the liver throughout the body, and when present in large amounts in the blood, it deposits and accumulates in blood vessel walls, causing atherosclerosis and resulting in myocardial infarction and cerebral infarction.

In humans, the normal range of LDL-cholesterol is less than 140 mg/dL, and more than 140 mg/dL is considered to reflect hyper-LDL cholesterolemia.

LabAssay™ LDL-Cholesterol is a kit for the determination of LDL-cholesterol in blood (serum and plasma). With the use of a microplate, this kit provides a quick and convenient method for measuring LDL-cholesterol in samples. In addition, our kit do not require fractionation of lipoproteins by centrifugation.

[Note] LabAssay™ series are reagents for research purposes. They cannot be used for diagnostic purposes.

Kit Performance

Animal Human, Mouse, Rat
Sample Serum, Plasma
Calibration curve range 9.38-300 mg/dL
Sample volume 5 μL
Measurement duration Approx. 20 min
Wavelength Primary wavelength 600 nm
Reference wavelength 700 nm

Example of Calibration Curve

04572869_img02.png

Principle

1) The first reaction(elimination of non-LDL cholesterol)

The block polymer containing hydrophilic and hydrophobic moiety in Pretreatment selectively binds to LDL in the specimens, and protects it from enzyme (CHE, CO) reactions. CHE and CO react with non-LDL lipoprotein [chylomicron (CM), very low density lipoprotein (VLDL), high density lipoprotein (HDL)]. Hydrogen peroxide produced by the enzyme reactions with non-LDL cholesterol is decomposed to water by catalase in Pretreatment reagent.

04572869_img03.png

2) The second reaction(color reaction of LDL-cholesterol)

When Reacting Solution is added, the cholesterol and its derivatives in LDL produce hydrogen peroxide by CHE and CO. Hydrogen peroxide produced by the enzyme reactions with LDL-cholesterol yields a blue color pigment upon oxidative condensation with N-(3-sulfopropyl)-3-methoxy-5-methylaniline (HMMPS) and 4-aminoantipyrine in the presence of peroxidase (POD). The amount of LDL-cholesterol contained in the sample is determined by measuring the absorbance of the blue color.

04572869_img04.png

Data

Precision of assay (within assay variation)

Reproducibility was confirmed by performing quintuplicate assays of serum samples of two concentrations each from humans, mice, and rats with this product.

n\ID Human Mouse Rat
ID1 (mg/dL) ID2 (mg/dL) ID1 (mg/dL) ID2 (mg/dL) ID1 (mg/dL) ID2 (mg/dL)
1 160 46.0 141 48.3 119 33.3
2 156 43.8 138 47.9 114 34.1
3 158 44.7 138 49.5 119 33.3
4 166 44.5 139 48.5 112 34.9
5 159 45.7 136 50.6 113 33.3
mean 160 44.9 138 49.0 115 33.8
SD 3.8 0.90 1.8 1.1 3.4 0.72
CV(%) 2.4 2.0 1.3 2.2 2.9 2.1

[Result]
The CV (%) of human serum was 2.0-2.4%, mouse serum was 1.3-2.2%, and rat serum was 2.1-2.9%, indicating good reproducibility.

Reproducibility (between assay variation)

Triplicate measurements were performed on human, mouse, and rat serum samples at three concentrations each for four days with this product.

Day\ID Human Mouse Rat
ID1 (mg/dL) ID2 (mg/dL) ID3 (mg/dL) ID1 (mg/dL) ID2 (mg/dL) ID3 (mg/dL) ID1 (mg/dL) ID2 (mg/dL) ID3 (mg/dL)
1 125 88.1 41.8 119 91.4 48.4 130 78.7 45.2
2 124 87.0 41.1 119 88.1 45.0 129 76.1 42.9
3 130 91.5 41.7 118 88.0 48.2 126 74.2 49.7
4 125 89.5 39.8 126 95.1 50.4 131 74.5 49.4
mean 126 89.0 41.1 121 90.7 48.0 129.0 75.9 46.8
SD 2.7 1.9 0.92 3.7 3.4 2.2 2.2 2.1 3.3
CV(%) 2.1 2.2 2.2 3.1 3.7 4.7 1.7 2.7 7.1

[Result]
The CV (%) of human serum was 2.1-2.2%, mouse serum was 3.1-4.7%, and rat serum was 1.7-7.1%, indicating good reproducibility.

Dilution linearity test

Using the standard diluent supplied with the product, human serum/plasma (EDTA), mouse serum/plasma (EDTA), and rat serum/plasma (EDTA) at two different concentrations each, were subjected to a 2-fold serial dilution with sample buffer and measured to confirm linearity (measured in duplicates).

04572869_img05.png
04572869_img06.png
04572869_img07.png
04572869_img08.png
04572869_img09.png
04572869_img10.png

[Result]
Good linearity was confirmed for all samples within the calibration curve range.

Spiked recovery test

Standard solutions at three concentrations were added to human serum/EDTA plasma, mouse serum/EDTA plasma and rat serum/EDTA plasma. Next, spike-recovery tests were conducted, with each measurement performed in duplicate.

Human

Amount spiked (mg/dL) Measured value (mg/dL) Recovery volume (mg/dL) Recovery rate (%)
Serum - 75.2 - -
7.12 82.3 7.10 100
28.5 103 27.8 97.5
35.6 112 36.8 103
Plasma
(EDTA)
- 84.5 - -
10.6 93.8 9.30 87.7
21.2 107 22.5 106
26.5 112 27.5 104

Mouse

Amount spiked (mg/dL) Measured value (mg/dL) Recovery volume (mg/dL) Recovery rate (%)
Serum - 86.6 - -
15.9 102 15.4 96.9
23.8 112 25.4 107
39.7 126 39.4 99.2
Plasma
(EDTA)
- 97.7 - -
14.4 112 14.3 99.3
28.9 126 28.3 97.9
36.1 133 35.3 97.8

Rat

Amount spiked (mg/dL) Measured value (mg/dL) Recovery volume (mg/dL) Recovery rate (%)
Serum - 14.6 - -
13.7 28.7 14.1 103
41.0 53.0 38.4 93.7
68.4 77.1 62.5 91.4
Plasma
(EDTA)
- 16.2 - -
16.0 31.9 15.7 98.1
48.1 64.4 48.2 100
80.2 90.6 74.4 92.8

[Result]
Good recovery rates was confirmed for all samples.

Measurement examples

Human serum/plasma, mouse (ICR) serum/plasma samples, and rat (SD) serum/plasma were measured with this product.

Human

No. Measured value (mg/dL) mean (mg/dL) SD CV(%)
Serum 1 189 187 188 1.4 0.75
2 88.8 93.2 91 3.1 3.4
3 85.7 85.7 85.7 0 0
4 99 93.9 96.5 3.6 3.7
5 95.8 94 94.9 1.3 1.3
Plasma
(EDTA)
1 122 121 122 0.71 0.58
2 88.3 88 88.2 0.21 0.24
3 85.9 84.4 85.2 1.1 1.2
4 93.9 97.2 95.6 2.3 2.4
5 95.1 95.4 95.3 0.21 0.22

Mouse

No. Measured value (mg/dL) mean (mg/dL) SD CV(%)
Serum 1 32.6 33.4 33 0.57 1.7
2 33.6 31.3 32.5 1.6 5
3 33.4 32.4 32.9 0.71 2.1
4 34.5 33.4 34 0.78 2.3
5 36.1 34.1 35.1 1.4 4
Plasma
(EDTA)
1 21.2 20.8 21 0.28 1.3
2 16 17.7 16.9 1.2 7.1
3 25.5 21.2 23.4 3 13
4 19.1 20.3 19.7 0.85 4.3
5 18.9 19.6 19.3 0.49 2.6

Rat

No. Measured value (mg/dL) mean (mg/dL) SD CV(%)
Serum 1 43.3 37.9 40.6 3.8 9.4
2 44.4 42.3 43.4 1.5 3.4
3 42.7 44 43.4 0.92 2.1
4 44.9 43.2 44.1 1.2 2.7
5 50 51.1 50.6 0.78 1.5
Plasma
(EDTA)
1 30.7 30.1 30.4 0.42 1.4
2 33.7 32.8 33.3 0.64 1.9
3 39.2 36.5 37.9 1.9 5
4 30.7 30.9 30.8 0.14 0.46
5 33.9 36.7 35.3 2 5.6

FAQ

About sample

What specimens should I use?
Use fresh specimens. Do not use specimens after repeated freeze-thawing as this may denature lipoproteins.
Which anticoagulants can I use?
Anticoagulants such as heparin, citrate and EDTA do not significantly influence the assay when used in normal amounts.
What should I do with samples that exceed the measurable range?
Dilute specimen with saline and repeat the assay if the measured value exceeds the measurable range, and multiply the result by the dilution factor.
How should I do when triglyceride in the specimen exceeds 1,000 mg/dL?
When triglyceride in the specimen exceeds 1,000 mg/dL, dilute the specimen with saline and multiply the result by the dilution factor.

About kit usage

What instruments, and equipment are required for the assay using this kit?
The instruments and equipment required for the use of this kit are listed below.
  • 96-well microplate (transparent type)
  • Micropipette
  • Incubator maintained at 37℃
  • Plate mixer
  • Microplate reader (with 600 nm/700 nm wavelength filter)
What is the amount of purified water to be added to the standard product?
Because the volume of distilled water to be added to the standard differs depending on the lot, see the specified volume in a separate sheet.

Overview / Applications

Property

Manufacturer Information

Alias

For research use or further manufacturing use only. Not for use in diagnostic procedures.

Product content may differ from the actual image due to minor specification changes etc.

If the revision of product standards and packaging standards has been made, there is a case where the actual product specifications and images are different.

+49-2131-311-271

Reception hours: Mon-Fri 9:00 - 15:00 (CET)For other hours than the above, please contact us via the inquiry form.