LBIS (TM) Human IL-8 (CXCL8) ELISA Kit

for Immunochemistry
Manufacturer :
FUJIFILM Wako Pure Chemical Corporation
Storage Condition :
Keep at 2-10 degrees C.
GHS :
  • Structural Formula
  • Label
  • Packing
SDS
Comparison
Product Number
Package Size
Price
Inventory
Distributor
296-87601
Barcode No
4548995102904
96Tests
List Price
JPY 78,000

In stock in Japan

Document

SDS
Product Specification Sheet
Package Insert
Spectral Data
Certificate of Analysis
Calibration Certificate
Analytical Charts

Kit component

96 tests

Antibody-coated Plate 96 wells/1 plate
Human IL-8 Standard 1 bottle
Buffer 60 mL/1 bottle
Biotin-conjugated Antibody 1 bottle
Peroxidase-conjugated Streptavidin Solution 100 μL/1 bottle
TMB Solution 12 mL/1 bottle
Stop Solution 12 mL/1 bottle
Wash Solution (10x) 100 mL/1 bottle
Plate Seal 4 sheets

Product Overview

Interleukin-8 (IL-8) - image

Interleukin-8 (IL-8) — also known as CXCL8 — is a chemokine produced by fibroblasts, monocytes, and vascular endothelial cells in response to stimuli such as inflammatory cytokines. Acting as a neutrophil chemoattractant, it functions as a key mediator of inflammation in the body. Its involvement has been studied in range of conditions including rheumatoid arthritis, respiratory diseases such as asthma, arteriosclerosis, gout, periodontitis, and cancer.

The LBIS™ Human IL-8 ELISA Kit is a sandwich ELISA kit designed for the quantitative measurement of human IL-8 (CXCL8). This kit enables highly sensitive quantification of IL-8 (CXCL8) in biological samples. The standard included in the kit is a recombinant protein produced in Escherichia coli.

Kit Performance

Assay target IL-8 (CXCL8)
Calibration curve range 0.686 – 500 pg/mL
Analysis sample Human Serum/Plasma (EDTA, heparin)/Cell culture supernatant*
Sample amount 50 μL (2x dilution)
Measurement duration Approx 3 hours and 50 minutes
Wavelength Primary wavelength : 450 nm
Reference wavelength : 620 nm
*Measurement availability depends on the culture medium. When using cell culture supernatant as a sample, please make sure to perform recovery test and dilution linearity test on the culture medium before use.

Example of Calibration Curve

Example of Calibration Curve

Data

Repeatability (within run precision)

Repeatability was conducted by measuring fivefold assays of human serum samples at two concentrations.

n\ID ID1 (pg/mL) ID2 (pg/mL)
1 169 42.4
2 162 41.7
3 165 44.9
4 155 43.7
5 179 45.1
mean 166 43.6
SD 8.9 1.5
CV(%) 5.4 3.4
[Result]
The CV (%) of human serum was 3.4-5.4%, indicating good repeatability.

Reproducibility (between run precision)

Reproducibility was conducted by measuring human serum samples at three concentrations each for four days.

Day\ID ID3 (pg/mL) ID4 (pg/mL) ID5 (pg/mL)
1 170 55.3 6.25
2 167 55.1 6.03
3 167 54.8 6.27
4 160 55.2 6.41
mean 166 55.1 6.24
SD 4.2 0.22 0.16
CV(%) 2.6 0.39 2.5
[Result]
The CV (%) of human serum was 0.39-2.6%, indicating good reproducibility.

Dilution Linearity Test

Human serum and plasma (EDTA/heparin), as well as cell culture supernatant spiked with the standard, were serially diluted twofold using the provided buffer and measured in duplicate to evaluate linearity.

Human serum

graph:Human serum

Human plasma (EDTA)

graph:Human plasma (EDTA)

Human plasma (heparin)

graph:Human plasma (heparin)

Human PBMC supernatants ± PHA stimulation (in RPMI-1640 medium with 10% FBS)

graph:Human PBMC supernatants ± PHA stimulation (in RPMI-1640 medium with 10% FBS)

Human PBMC supernatants stimulated with/without PHA (phytohemagglutinin, simulating IL-8 secretion) were measured at 1:500 / 1:50 dilution.

[Result]
All measurement results showed good linearity within the measurement range.

Spike and Recovery Test

Standard solutions at three concentrations were added to human serum/plasma (EDTA), then spike recovery tests were conducted, with each measurement performed in duplicate.

Human serum

Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery amount
(pg/mL)
Recovery
(%)
0 18.3 - -
62.5 78.6 60.3 96.5
125 145 127 101
250 248 230 91.9
mean 96.6

Human plasma (EDTA)

Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery amount
(pg/mL)
Recovery
(%)
0 1.29 - -
9.38 11.0 9.7 104
32.5 32.2 30.9 95.1
130 130 129 99.0
mean 99.2

Human plasma (heparin)

Spiked amount
(pg/mL)
Measurement value
(pg/mL)
Recovery amount
(pg/mL)
Recovery
(%)
0 4.01 - -
15.4 19.2 15.2 98.6
62.5 70.1 66.1 106
125 125 121 96.8
mean 100
[Result]
Good recovery rates were confirmed.

Example of Sample Measurement

Human serum and plasma (EDTA/heparin) samples were measured and compared using this kit and the competitor.

Human serum

ID\n Fujifilm Wako
(pg/mL)
Company A
(pg/mL)
1 1.90 N.D.
2 10.5 N.D.
3 12.8 N.D.
4 19.4 21.6
5 36.9 42.9
6 0.864 N.D.
7 16.4 5.77
8 80.9 141
9 9.47 N.D.
10 7.08 N.D.

Human plasma (EDTA)

ID\n Fujifilm Wako
(pg/mL)
Company A
(pg/mL)
11 1.94 N.D.
12 15.7 N.D.
13 3.69 N.D.
14 3.19 N.D.
15 3.36 N.D.
16 5.36 N.D.
17 3.85 N.D.
18 3.36 N.D.
19 1.09 N.D.
20 0.874 N.D.

Human plasma (heparin)

ID\n Fujifilm Wako
(pg/mL)
Company A
(pg/mL)
21 2.20 N.D.
22 4.57 N.D.
23 5.09 N.D.
24 23.1 25.7
25 3.14 N.D.
26 2.33 N.D.
27 2.17 N.D.
28 1.30 N.D.
29 3.03 N.D.
30 3.70 N.D.

N.D.: Below the lower limit of quantification

[Result]
This product was able to detect IL-8 (CXCL8) even in samples that were below the lower limit of quantification of another manufacturer's kit.

Correlation with conventional product

For mouse serum samples, measurements were conducted using this kit and competitor product B (high sensitivity product). The measurement values obtained using both products were then compared.

Correlation with conventional product
[Result]
Good correlation was confirmed.

FAQ

About samples

What anticoagulants should I use?
We recommend using EDTA or heparin as anticoagulant.
How should I store and prepare the samples?
Samples should be immediately assayed or stored below -35℃ until assay. Before starting assay, shake thawed samples sufficiently. Do not repeat freeze-and-thaw cycles. Dilution of a sample should be made in a PP or PE test tube using the provided buffer prior to adding them to wells.

About kit usage

What instruments, and equipment are required for the assay using this kit?
The instruments and equipment required for the use of this kit are listed below;
  • Deionized water (or distilled water)
  • Test tubes for preparation of standard solution series
  • Glassware for dilution of Wash Solution (10x) (a graduated cylinder, a bottle)
  • Pipettes (disposable tip type). One should be able to deliver 50 μL precisely, and another for 100 – 1,000 μL
  • Syringe-type repeating dispenser like Eppendorf multipette plus which can dispense 100 μL
  • Paper towel to remove washing buffer remaining in wells
  • A vortex-type mixer
  • A shaker for 96 well-plate (600 – 1,200 rpm)
  • An automatic washer for 96 well-plate (if available), or a wash bottle with a jet nozzle
  • A 96 well-plate reader (450 ± 10 nm, 620 nm :600 - 650 nm)
  • Software for data analysis
How should I use the human IL-8 standard?
The amount of purified water to be added varies by lot. Please reconstitute the human IL-8 standard with the amount of purified water specified in the “Reconstitution of standard” section on this product page to prepare the original standard solution (3 ng/mL). Then, dilute the solution using the kit-provided buffer that has been equilibrated to room temperature.
What are the storage conditions for each reagent?
[Buffer & TMB Solution]
Use only volume you need for your assay. Remaining reagents should be stored at 2-10℃ closing the cap tightly. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.
[Stop Solution]
Close the cap tightly and store at 2-10℃.
[Human IL-8 Standard & Biotin-conjugated Antibody]
The reconstituted standard solution (original standard solution)  should be used within 2 weeks if stored in a refrigerator. Dispose the remaining diluted standard solutions after use.
[Peroxidase-conjugated Streptavidin Solution]
Remaining working solution (already diluted) should be disposed. The rest of the undiluted solution (unused): close the cap tightly and store at 2-10℃. Once opened, we recommend using as soon as possible to avoid influence by environmental condition.

Prepare reagent solutions in appropriate volume for your assay.

Overview / Applications

Property

Manufacturer Information

Alias

  • 632-42321 (Previous code)
    AKH-IL8 (Previous code)

For research use or further manufacturing use only. Not for use in diagnostic procedures.